background image

36

ISE SETUP (HI 5222 only)

The ISE Setup menu allows the user to set the parameters regarding ISE measurement and calibration. These
parameters can be set specifically for each channel. The settings will be applied only to the active channel.

Accessing ISE Setup

• Press 

 while in 

Measure

 mode and then 

 to

select ISE range for the desired channel.

• Press 

 and then

 

 to access ISE Setup menu.

To access an ISE Setup option:

• Use 

 or 

 to highlight the desired option.

• Press 

 to access the selected option.

The following is a detailed description of the ISE Setup option
screens.

Profile

 

- See pH Setup section.

Reading Mode

This option allows the user to select the desired reading mode: Direct, Direct/AutoHold, Known Addition, Known
Subtraction, Analyte Addition and Analyte Subtraction. Four of these Reading Modes are collectively known as
Incremental Methods (see ISE Theory section for details). Direct measurements and Direct/AutoHold measurements
are also available.

Direct

Direct measurements are analogous to taking pH measurements. The ISE is calibrated in Ion standards and
sample measurements are made directly. The ISE’s manual should be consulted for tips and practices of making
Direct measurements. The Ion concentration can be read directly from the instrument.

Direct/AutoHold

Direct/AutoHold measurements are made similar to Direct measurements. The advantage of using AutoHold is a
measurement that has not reached equilibrium will not be used. Only after the chosen stability criteria has been
met will the meter go into the AutoHold mode. Using AutoHold removes the subjective nature of stability.

Known Addition

In the Known Addition method, a sample is measured with an ISE before and after the addition of a known volume
of a standard. The mV difference is then used to calculate the concentration of the Ion in the original sample.

45

• Press 

. If the instrument was calibrated before and calibration was not cleared, the old calibration

can be cleared by pressing 

. After 10 seconds, 

 will no longer be available.

Note:

It is very important to clear calibration history when a new electrode is used because most errors
and warning messages that appear during calibration depend on calibration history.

• Immerse the pH electrode and the temperature probe approximately 4 cm (1.5”) into

a buffer solution of your choice (pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01,
12.45 or a custom buffer) and stir gently. The temperature probe should be close to
the pH electrode.

• Select the pH calibration buffer used with 

 or 

. The “

Please wait...

message will appear on the LCD until the reading is stable or the buffer is validated.

• If the pH buffer is validated, 

 will appear on the LCD. Press 

 to update calibration. The

calibration buffer will be added to the Calibrated Buffers section.

• Immerse the pH electrode and the temperature probe into the next buffer solution and follow the above

procedure or press 

 to exit calibration.

Notes:

• The new added calibration point will replace an old

one if the difference between them is ±0.2 pH.

• If the existing stored calibration is full (five calibration

points), a pop-up menu will be displayed on the LCD
in which you can select with 

 or 

keys the buffer you want to replace with the current
buffer. Press 

 to delete the selected buffer

and then press 

 to update calibration with the

new buffer.

pH Calibration screen description

Содержание HI 5221

Страница 1: ...i n s t c o m MAN5222 02 15 PrintedinROMANIA Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 757...

Страница 2: ...e mail us at tech hannainst com or view our worldwide contact list at www hannainst com WARRANTY All rights are reserved Reproduction in whole or in part is prohibited without the written consent of t...

Страница 3: ...king material and examine it carefully to make sure that no damage has occurred during shipping If there is any damage notify your Dealer or the nearest Hanna Customer Service Center PRELIMINARY EXAMI...

Страница 4: ...HI 5222 same for each measurement channel Log files include GLP information Basic Measurement can be viewed with detailed GLP information or with a Graph or a Log History User friendly interface on la...

Страница 5: ...age Profile feature store up to ten different user setup HI 5222 five on each channel 76 HI 2031B Glass body semimicro conical refillable combination pH electrode Use semisolid products HI 1332B Plast...

Страница 6: ...nnel 2 10 Temperature probe socket Channel 2 11 BNC electrode connector for pH ORP ISE measurements Channel 2 REAR PANEL 75 HI 1131B Glass body refillable double junction combination pH electrode Use...

Страница 7: ...eral information about the selected option operation 74 ORP SOLUTIONS HI 7020L Test Solution 200 275 mV 500 mL bottle HI 7021L Test Solution 240 mV 500 mL bottle HI 7022L Test Solution 470 mV 500 mL b...

Страница 8: ...p m I t u p n I 0 1 2 1 s m h o y a l p s i D s l e x i p 0 2 3 x 0 4 2 D C L c i h p a r G r o l o C y l p p u S r e w o P r e t p a d a c d V 2 1 t n e m o r i v n E g n i s n e d n o c n o n H R 5...

Страница 9: ...utton located on the rear panel of the instrument Please wait until the instrument finishes the initialization process Note It is normal for the loading process to take a few seconds If the instrument...

Страница 10: ...n i e r a n o i t c n u j e c n e r e f e r e h t n i s i e l p m a s e h t e r u s e k a M e h t k c e h C e g n a r d e i f i c e p s l a r e n e g e h t d n a l e v e l e t y l o r t c e l e E S I...

Страница 11: ...r a faster response and to avoid cross contamination of the samples rinse the electrode tip with a few drops of the solution to be tested before taking measurements STORAGE PROCEDURE To minimize clogg...

Страница 12: ...he protective cap off the pH electrode SALT DEPOSITS MAY BE PRESENT This is normal with electrodes They will disappear when rinsed with water During transport tiny bubbles of air may form inside the g...

Страница 13: ...emperature During calibration the instrument will display the pH buffer value at 25 C P M E T S R E F F U B H p C K F 9 7 6 1 0 0 0 3 0 1 0 4 2 6 8 6 0 1 0 7 7 7 1 9 0 1 0 0 1 4 5 4 2 1 0 3 7 2 2 3 0...

Страница 14: ...MP the sample concentration CSTD the standard concentration VSAMP the sample volume VSTD the standard volume VISA ISA volume E the difference of potential from the electrode S the electrode slope dete...

Страница 15: ...adjustment is made to samples and standards Unknowns are measured directly by the instrument At lower concentrations in non linear regions of the electrode response multiple calibration points will e...

Страница 16: ...e Press 65 ADDITIONAL INFORMATION ISE THEORY An Ion Selective Electrode ISE is an electrochemical sensor that changes voltage with the activity or concentration of ions in solutions The change in volt...

Страница 17: ...ion Press and use key to highlight the desired parameter Use or to adjust the contrast backlight or to set the desired backlight saver time Press to confirm the modified options and return to the Syst...

Страница 18: ...ss and then the desired parameter or HI 5222 only Only the selected measurement parameter lots will be displayed on the LCD Select the desired lot with or and press to display the logged data from the...

Страница 19: ...meter reaches the stability criteria to go into Auto Hold mode The stored sample logs will be marked with an H to indicate the Auto Hold mode The virtual key returns operation to real time continuous...

Страница 20: ...e r i D 2 l a u n a M g o L l a u n a M A N d l o H o t u A e l b a c i l p p A t o N 3 c i t a m o t u A g o L c i t a m o t u A d l o H o t u A t c e r i D 4 l a u n a M g o L l a u n a M 5 d l o H...

Страница 21: ...p of the meter and ensures that the same procedure will be used To save the measurement configuration for pH mode Press and use or to highlight Profile Press to enable disable this feature pH SETUP 60...

Страница 22: ...elect the Analyte Subtraction method see ISE Setup Reading Mode Prior to starting an AS procedure the ISE sensor must be calibrated with a minimum of two standards containing ISA The slope of the elec...

Страница 23: ...he best results As a general rule the standard must be less concentrated than the sample so the addition of sample will increase the total Ion content of the beaker and change the mV value by at least...

Страница 24: ...the observed mV Using a micropipet add a volume of the Reactant standard Start by adding 500 L at a time for example Watch the change in mV When you have observed approximately a 15 mV change from the...

Страница 25: ...the solution and a mV value will show on the display Press to take the first mV reading When the reading is stable press to store the first mV reading The second step of the method will be displayed o...

Страница 26: ...od Press and then to select ISE Measure mode for the selected channel Select the Known Addition method see ISE Setup for details Prior to starting a KA procedure the ISE sensor must be calibrated with...

Страница 27: ...e of the incremental methods for measurement at least a two point ISE calibration must be performed to establish the electrode slope For accurate measurements add the appropriate ISA Ionic Strength Ad...

Страница 28: ...oosing the AutoHold option the current reading can be frozen on the LCD when is pressed and the stability criterion is reached 53 If the isopotential point of the electrode is unknown the ISE calibrat...

Страница 29: ...ared by pressing After 10 seconds will no longer be available Note It is very important to clear calibration history when a new electrode is used because most errors and warning messages that appear d...

Страница 30: ...BRATION HI 5222 only For greater accuracy it is recommended to calibrate the ISE sensors frequently The instrument should be recalibrated when ISE x Calibration Expired the x represents channel 1 or c...

Страница 31: ...larm State Press while in pH Measure mode Press Use or to select the Alarm option Press anduse or tohighlighttheAlarm State option Press anduse or tohighlightthedesired option Press to confirm your se...

Страница 32: ...UREMENTS mV ORP MEASUREMENTS Oxidation reduction potential ORP measurements provide the quantification of the oxidizing or reducing power of the tested sample To correctly perform a redox measurement...

Страница 33: ...le solution The pH reading is affected by temperature In order to measure the pH accurately the temperature effect must be compensated To use the Automatic Temperature Compensation ATC feature connect...

Страница 34: ...for HI 5222 Select the Direct reading mode see pH Setup for details Place the electrode tip and the temperature probe approximately 4 cm 1 5 into the sample to be tested Allow time for the electrode...

Страница 35: ...value can be adjusted using or Press to save the new temperature value If using Custom buffers press after buffer has been accepted to change actual buffer conditions A pop up menu will be displayed o...

Страница 36: ...ctive nature of stability Known Addition In the Known Addition method a sample is measured with an ISE before and after the addition of a known volume of a standard The mV difference is then used to c...

Страница 37: ...trode Cond Unknown pH x Default Calibration or pH x Calibration Expired message appears on the LCD in the Reminder messages area The x represents channel 1 or channel 2 PREPARATION Pour small quantiti...

Страница 38: ...benefit from temperature compensated corrections if standards and sample temperatures differ from each other the Isopotential Point of the ISE is known If sample and standards are made at the same tem...

Страница 39: ...key to select the next custom standard to be set Press to return to Edit Custom Standard options 42 To select the appropriate Ion Charge Slope use or and then press If the Ion electric charge is None...

Страница 40: ...n constants Name Molar Weight and Electric Charge Slope while for the custom ISE all these constants can be manually set 41 To set the Electrode Type Press while in ISE Measure mode Press Use or to se...

Отзывы: