17
• Use
or
to select the Date & Time option.
• Press
and use
or
to highlight the
desired option you want to modify.
• Press
to confirm your selection. Use
/
to select next/previous entry to be edited. Press
and
use
or
to set the desired value, then press
to save the modified value (for Set Date and Time
option). For the Set Date Format option press
to confirm
your selection and select one of the displayed options with
or
.
• Press
to confirm your selection and return to the Date & Time options or press
to cancel
operation.
Note
: If the time is changed with more than one hour before last pH/ISE calibration, a pop-up warning
will appear on the LCD, notifying the user that a date/time conflict has occurred and some
time-dependent modes could work improperly (e.g.
Measure
, GLP, Log).
LCD Setup
This option allows the user to set the
Contrast
, the
Backlight
of the LCD and the
Backlight Saver
. The Contrast
parameter can be adjusted within 7 steps, while the Backlight parameter within 8 steps. The Backlight Saver can
be set from 1 to 60 minutes or it can be OFF (disabled). All the changes are visible on the LCD for each parameter.
Note
: If the instrument backlight turns off after the set period of time, press any key to turn it back on.
To set the LCD Setup:
• Press
while in
Measure
mode.
• Press
.
• Use
or
to select the LCD Setup option.
• Press
and use
key to highlight the desired
parameter.
• Use
or
to adjust the contrast / backlight or to
set the desired backlight saver time.
• Press
to confirm the modified options and return to the
System Setup menu.
64
PC INTERFACE
Data transmission from the instrument to the PC can be done with the
HI 92000
Windows
®
compatible
software (optional).
HI 92000
also offers graphing and on-line help features.
Data logged on the
HI 5221
and
HI 5222
meters can be exported to the most popular spreadsheet programs
for further analysis.
HI 5221
and
HI 5222
instruments have an USB interface available.
Use a standard USB cable to connect your instrument to the PC.
Make sure that the instrument and the
HI 92000
software have the same baud rate and the appropriate
communication port.
The PC software may also be used for real time logging.
To delete lots
:
• Press
while in
Log Recall
mode.
• Press
or
to access delete or delete all
mode. Otherwise, press
to return to
Log Recall
view
mode.
• After selecting one of the deleting modes, use
or
to select one lot and then press
or
to delete the selected lot or all lots. The “
Please wait...
”
message will be displayed on the LCD until the selected lot or
all lots are deleted.
• Press
and then press
to exit deleting mode
and return to
Log Recall
view mode.
• Press
to exit
Log Recall
mode and return to
Measure
mode.
Note
: Logged lots should also be deleted whenever “
Please Delete Old Log Files
” or “
Low Data Logging
Space
” message appears on the LCD, in the Reminder messages area.