To generate a New Lot:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the Log option.
• Press
and use
or
to highlight the New
Lot option.
• Press
to generate a new manual lot. A pop-up menu will
be displayed asking for confirmation.
• Press
to confirm or press
to escape without
saving and return to the Log options.
Alarm
This option allows the user to select the alarm settings: Alarm State and Alarm Limits. If the Alarm option
is enabled, a continuous double beep will be heard, along with the “
Alarm
” indicator blinking on the LCD,
each time the set limits in Measure mode are exceeded.
Note:
Alarm Beeper must be set On for audible beep to be heard.
See: System Setup > Beeper > Alarm.
Alarm State
Three settings are available for the Alarm State option:
Disabled
- the alarm will be disabled.
Inside Limits
- the alarm state will trigger when the measured value is inside the set limits.
Outside Limits
- the alarm state will trigger when the measured value is outside the set limits.
To set the Alarm State:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the Alarm option.
• Press
and use
or
to highlight the Alarm
State option.
• Press
and use
or
to highlight the desired
option.
• Press
to confirm your selection or press
to cancel
operation.
Alarm Limits
This option allows the user to set the alarm limits for the measured value.
Note
: The Alarm High value can not be lower than the Alarm Low value.
50
Relative mV MEASUREMENTS
To measure the Relative mV of a sample:
• Press
then
(select channel first for
HI 5222
).
• Verify if a current calibration has been made.
• If required, conduct the single point Rel mV calibration.
Verify the tip of the electrode is immersed into the known
solution or ORP standard.
• Press
. Use
and
arrow keys set the
standard value. Press
to store the calibration.
• Press
then
(select channel first for
HI 5222
).
• Place calibrated sensor tip into the sample to be analyzed. The
instrument will display the measured Relative mV value on the
LCD, together with a short GLP information about the last
calibration or Offset: 0.0 mV no Rel mV calibration was
performed.
Notes:
If the ORP sensor is not in solution or the measured mV
potential is out of range, “
----
”
will be displayed on the
LCD.