background image

23

To set one of the Temperature options:
• Press 

 while in 

pH Measure

 mode.

• Press 

.

• Use 

 or 

 to select the Temperature option.

• Press 

 and use 

 or 

 to highlight the

desired Temperature option you wish to modify.

• Press 

 and use 

 or 

 to highlight the

desired option (for Temperature Source & Unit options) or use

 or 

 to adjust the temperature value between

the displayed limits (for Manual Temperature option).

• Press 

 to confirm your selection (for Temperature Source

& Unit options) or press 

 to save the current value (for Manual Temperature option). Otherwise, press

 

to cancel operation.

Calibration

This option allows the user to setup desired buffers and calibration reminder.

Buffer Entry Type

Three settings are available for the pH buffers used for electrode calibration:

Automatic

 - the instrument automatically selects the closest buffer to the measured pH value from the

predefined buffers chosen in the option 

Edit Buffer Group

.

Semiautomatic

 - the instrument automatically selects the closest buffers to the measured pH value from all

available buffers and you can choose the one used to be, from standard and custom buffers.

Manual Selection

 - the desired pH buffer is manually selected from all available buffers (standard and

custom).

To set the Buffer Entry Type:
• Press 

 while in 

pH Measure

 mode.

• Press 

.

• Use 

 or 

 to select the Calibration option.

• Press 

 and use 

 or 

 to highlight the

Buffer Entry Type option.

• Press 

 and use 

 or 

 to highlight the

desired option.

• Press 

 to confirm your selection or press 

 

to

cancel operation.

58

• Press 

 to set the method parameters. Press 

 / 

 to select next/previous parameter

to edit, then press 

 and use 

 or 

 to set the desired parameter value. Press

 to save the modified value and then press 

 to exit method parameters edit menu.

• If developing a procedure: Before attempting Analyte Addition analysis, it is important to determine

which standard volume, concentration and sample size will produce the best results. As a general rule,
the standard must be less concentrated than the sample so the addition of sample will increase the total
Ion content of the beaker and change the mV value by at least 10 mV. For a positively charged Ion (i.e.
Sodium), the AA increases the mV. For a negatively charged Ion (i.e. Sulfide, Fluoride, Chloride), the AA
should decrease the mV. Start with a small trial. For example: Measure 50 mL of standard, add a
magnetic stir bar and place on a stirrer, add ISA (consult ISE manual) and place ISE electrode tip into the
sample. Put instrument in mV mode and record the observed mV. Using a micropipet, add a volume of
the sample. Start by adding 500 µL at a time (for example). Watch the change in mV. When you have
observed approximately a 10 mV change from the original standard, calculate the total volume added.
Adjust sample and standard volumes proportionally to standard volumes that can be measured with
accuracy. Use volumetric pipettes for standard, ISA and sample addition.

• Press 

 then edit the procedure variables to the volumes determined in the prior step.

Procedure:

• Press 

 to enter Analyte Addition mode.

• Volumetrically add standard to a clean beaker. Add Stir bar and place on a magnetic stir plate. Stir

standard. The method will prompt user to add ISA. Place ISE sensor tip into the solution and a mV value
will show on the display.

• Press 

 to take the first mV reading.

• When the reading is stable, press 

 to store the first mV reading. The second step of the method

will be displayed on the LCD, in which the user is notified to add the Sample Volume to the standard
solution. The method parameters are also displayed on the LCD.

• Press 

 to take the second mV reading.

• When the reading is stable, press 

 to store the second mV reading. The ISE measurement results

will be displayed on the LCD.

• Press 

 to log the current results into an ISE Method Report. Press 

 to return to 

ISE

Measure

 mode.

• Press 

 to start another measurement. Rinse ISE sensor between samples.

• Press 

, to modify the method parameters.

Note

: Press 

 at any time to stop the measurement and return to 

ISE Measure

 mode.

Содержание HI 5221

Страница 1: ...i n s t c o m MAN5222 02 15 PrintedinROMANIA Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 757...

Страница 2: ...e mail us at tech hannainst com or view our worldwide contact list at www hannainst com WARRANTY All rights are reserved Reproduction in whole or in part is prohibited without the written consent of t...

Страница 3: ...king material and examine it carefully to make sure that no damage has occurred during shipping If there is any damage notify your Dealer or the nearest Hanna Customer Service Center PRELIMINARY EXAMI...

Страница 4: ...HI 5222 same for each measurement channel Log files include GLP information Basic Measurement can be viewed with detailed GLP information or with a Graph or a Log History User friendly interface on la...

Страница 5: ...age Profile feature store up to ten different user setup HI 5222 five on each channel 76 HI 2031B Glass body semimicro conical refillable combination pH electrode Use semisolid products HI 1332B Plast...

Страница 6: ...nnel 2 10 Temperature probe socket Channel 2 11 BNC electrode connector for pH ORP ISE measurements Channel 2 REAR PANEL 75 HI 1131B Glass body refillable double junction combination pH electrode Use...

Страница 7: ...eral information about the selected option operation 74 ORP SOLUTIONS HI 7020L Test Solution 200 275 mV 500 mL bottle HI 7021L Test Solution 240 mV 500 mL bottle HI 7022L Test Solution 470 mV 500 mL b...

Страница 8: ...p m I t u p n I 0 1 2 1 s m h o y a l p s i D s l e x i p 0 2 3 x 0 4 2 D C L c i h p a r G r o l o C y l p p u S r e w o P r e t p a d a c d V 2 1 t n e m o r i v n E g n i s n e d n o c n o n H R 5...

Страница 9: ...utton located on the rear panel of the instrument Please wait until the instrument finishes the initialization process Note It is normal for the loading process to take a few seconds If the instrument...

Страница 10: ...n i e r a n o i t c n u j e c n e r e f e r e h t n i s i e l p m a s e h t e r u s e k a M e h t k c e h C e g n a r d e i f i c e p s l a r e n e g e h t d n a l e v e l e t y l o r t c e l e E S I...

Страница 11: ...r a faster response and to avoid cross contamination of the samples rinse the electrode tip with a few drops of the solution to be tested before taking measurements STORAGE PROCEDURE To minimize clogg...

Страница 12: ...he protective cap off the pH electrode SALT DEPOSITS MAY BE PRESENT This is normal with electrodes They will disappear when rinsed with water During transport tiny bubbles of air may form inside the g...

Страница 13: ...emperature During calibration the instrument will display the pH buffer value at 25 C P M E T S R E F F U B H p C K F 9 7 6 1 0 0 0 3 0 1 0 4 2 6 8 6 0 1 0 7 7 7 1 9 0 1 0 0 1 4 5 4 2 1 0 3 7 2 2 3 0...

Страница 14: ...MP the sample concentration CSTD the standard concentration VSAMP the sample volume VSTD the standard volume VISA ISA volume E the difference of potential from the electrode S the electrode slope dete...

Страница 15: ...adjustment is made to samples and standards Unknowns are measured directly by the instrument At lower concentrations in non linear regions of the electrode response multiple calibration points will e...

Страница 16: ...e Press 65 ADDITIONAL INFORMATION ISE THEORY An Ion Selective Electrode ISE is an electrochemical sensor that changes voltage with the activity or concentration of ions in solutions The change in volt...

Страница 17: ...ion Press and use key to highlight the desired parameter Use or to adjust the contrast backlight or to set the desired backlight saver time Press to confirm the modified options and return to the Syst...

Страница 18: ...ss and then the desired parameter or HI 5222 only Only the selected measurement parameter lots will be displayed on the LCD Select the desired lot with or and press to display the logged data from the...

Страница 19: ...meter reaches the stability criteria to go into Auto Hold mode The stored sample logs will be marked with an H to indicate the Auto Hold mode The virtual key returns operation to real time continuous...

Страница 20: ...e r i D 2 l a u n a M g o L l a u n a M A N d l o H o t u A e l b a c i l p p A t o N 3 c i t a m o t u A g o L c i t a m o t u A d l o H o t u A t c e r i D 4 l a u n a M g o L l a u n a M 5 d l o H...

Страница 21: ...p of the meter and ensures that the same procedure will be used To save the measurement configuration for pH mode Press and use or to highlight Profile Press to enable disable this feature pH SETUP 60...

Страница 22: ...elect the Analyte Subtraction method see ISE Setup Reading Mode Prior to starting an AS procedure the ISE sensor must be calibrated with a minimum of two standards containing ISA The slope of the elec...

Страница 23: ...he best results As a general rule the standard must be less concentrated than the sample so the addition of sample will increase the total Ion content of the beaker and change the mV value by at least...

Страница 24: ...the observed mV Using a micropipet add a volume of the Reactant standard Start by adding 500 L at a time for example Watch the change in mV When you have observed approximately a 15 mV change from the...

Страница 25: ...the solution and a mV value will show on the display Press to take the first mV reading When the reading is stable press to store the first mV reading The second step of the method will be displayed o...

Страница 26: ...od Press and then to select ISE Measure mode for the selected channel Select the Known Addition method see ISE Setup for details Prior to starting a KA procedure the ISE sensor must be calibrated with...

Страница 27: ...e of the incremental methods for measurement at least a two point ISE calibration must be performed to establish the electrode slope For accurate measurements add the appropriate ISA Ionic Strength Ad...

Страница 28: ...oosing the AutoHold option the current reading can be frozen on the LCD when is pressed and the stability criterion is reached 53 If the isopotential point of the electrode is unknown the ISE calibrat...

Страница 29: ...ared by pressing After 10 seconds will no longer be available Note It is very important to clear calibration history when a new electrode is used because most errors and warning messages that appear d...

Страница 30: ...BRATION HI 5222 only For greater accuracy it is recommended to calibrate the ISE sensors frequently The instrument should be recalibrated when ISE x Calibration Expired the x represents channel 1 or c...

Страница 31: ...larm State Press while in pH Measure mode Press Use or to select the Alarm option Press anduse or tohighlighttheAlarm State option Press anduse or tohighlightthedesired option Press to confirm your se...

Страница 32: ...UREMENTS mV ORP MEASUREMENTS Oxidation reduction potential ORP measurements provide the quantification of the oxidizing or reducing power of the tested sample To correctly perform a redox measurement...

Страница 33: ...le solution The pH reading is affected by temperature In order to measure the pH accurately the temperature effect must be compensated To use the Automatic Temperature Compensation ATC feature connect...

Страница 34: ...for HI 5222 Select the Direct reading mode see pH Setup for details Place the electrode tip and the temperature probe approximately 4 cm 1 5 into the sample to be tested Allow time for the electrode...

Страница 35: ...value can be adjusted using or Press to save the new temperature value If using Custom buffers press after buffer has been accepted to change actual buffer conditions A pop up menu will be displayed o...

Страница 36: ...ctive nature of stability Known Addition In the Known Addition method a sample is measured with an ISE before and after the addition of a known volume of a standard The mV difference is then used to c...

Страница 37: ...trode Cond Unknown pH x Default Calibration or pH x Calibration Expired message appears on the LCD in the Reminder messages area The x represents channel 1 or channel 2 PREPARATION Pour small quantiti...

Страница 38: ...benefit from temperature compensated corrections if standards and sample temperatures differ from each other the Isopotential Point of the ISE is known If sample and standards are made at the same tem...

Страница 39: ...key to select the next custom standard to be set Press to return to Edit Custom Standard options 42 To select the appropriate Ion Charge Slope use or and then press If the Ion electric charge is None...

Страница 40: ...n constants Name Molar Weight and Electric Charge Slope while for the custom ISE all these constants can be manually set 41 To set the Electrode Type Press while in ISE Measure mode Press Use or to se...

Отзывы: