29
SETUP
14. SETUP
Setup mode allows viewing and modifying the instrument parameters.
Note:
Blinking CAL tag during setup mode indicates to press CAL for parameters editing.
1. Press SETUP/DEL key to enter or exit Setup mode.
2. Press GLP
/ AVG
keys to select the desired panel with parameter to be edited.
Beep
Options: ON or OFF
This parameter indicates error conditions, a wrong key press or a tag read.
1. Press CAL when beep panel is displayed. Status and CFM tag are displayed blinking.
2. Press GLP
/ AVG
keys to set desired option.
3. Press LOG/CFM to save.
Alternatively, press CAL to exit without saving.
Time (Hour & Minutes)
Options: Show or Hide
1. Press CAL when show/hide time panel is displayed. Status and CFM tag are displayed blinking.
2. Press GLP
/ AVG
keys to set desired option.
3. Press LOG/CFM to save.
Alternatively, press CAL to exit without saving the changes.
Date
Options: year, month, day
The instrument’ s internal RTC clock is used to generate a unique time stamp for each recorded value
and to automatically store the last calibration date. The current time can be displayed on the LCD
when the instrument is in idle mode.
1. Press CAL in set date panel. Date format is YYYY.MM.DD. Selected value and CFM tag is displayed
blinking.
2. Press LOG/CFM or READ
to select between options. Selection blinks.