LP 2000

 displays several different codes and messages to help the

user to operate the meter more efficiently.

This indicates that the meter is in a ready
state and measurement or calibration can be
performed.

This indicates that the calibration mode is
active. If the CAL key is not pressed within 6
seconds, the meter will automatically revert
back to the measurement mode.

This indicates that the calibration date setting
mode is active.

This indicates that a 0 FTU standard calibra-
tion solution is required.

This indicates that a 10 FTU standard calibra-
tion solution is required.

This indicates that a 500 FTU standard cali-
bration solution is required.

This blinking SIP (Sample In Progress) mes-
sage indicates that the instrument is being
calibrated.

This shows the last calibration date (MM.DD).

This indicates that the sample is being ana-
lyzed.

This indicates that an error occurred during
calibration (e.g. a wrong standard was used).
Repeat the calibration with fresh standards.

LCD AND ERROR CODES

TIPS TO OBTAIN ACCURATE CALIBRATION

All glassware that comes into contact with standards should be
kept clean. Wash with 

HI 93703-50 

cleaning solution and rinse

with 

HI 93703-0 

or turbidity free water.

Rinse the cuvette twice with 5 mL of the sample, to remove any
previous liquid, dust or foreign matter that may be present.
Gently pour the liquid down the side of the vial to reduce air
bubbles (no mixing is required when Hanna standards are used).

Before inserting the vial into the meter,
wipe it clean with 

HI93703-70 

or a soft,

lint-free tissue. Handle vials so that no
fingerprints are left on the areas where
light passes through (bottom 2 cm).

STANDARD SUSPENSION

Presently, there are only two recognized primary standards: AMCO-AEPA-1
(available through Hanna) and formazine.

 LP 2000 

works with both

solutions.
AMCO-AEPA-1 has a much longer shelf life at all concentrations
(approximately six months, if free from contamination). In addition,
no special handling or disposal is required and a much higher
stability of suspended particles has been observed.
Formazine is highly toxic and is generated by a known carcinogen,
with poor stability (particles settle quickly). Lower concentrations
change value quickly after dilution.
The consistency of 

LP 2000 

readings by using both standards has

been separately established by Advanced Polymer Systems and
Hanna Instruments. Additional documentation about the formazine
standard and other calibration procedures is available upon request.

VIEWING THE CALIBRATION DATE

Turn the meter on and wait for the
display to show "----".

Press and hold the DATE/   key. The month and day of last
calibration ("MM.DD") appear while the key is held down.

Note:

The displayed date is the date that was input by the user at
the beginning of the last calibration.

Содержание LP 2000

Страница 1: ...Instruction Manual LP 2000 Precision Bench Turbidity Meter w w w h a n n a i n s t c o m ...

Страница 2: ...LCD The meter is simple to use all operations can be carried out with only four keys and troubleshooting functions can be performed with error codes on the display An exclusive positive locking system guarantees that the cuvette is firmly placed in the cell The keypad is water resistant and can be wiped clean Calibration at 10 FTU can be easily performed using the supplied standard In addition LP ...

Страница 3: ...1 19 2 5 FTU NTU 0 053 1 0 13 SiO2 mg L 0 4 7 5 1 MEASURE UNITS LP 2000 has been designed to perform measurements according to the ISO 7027 International Standards The instrument functions by passing a beam of infrared light through a vial containing the sample being measured The light source is a High Emission Infrared LED with a wavelength peaking at 890 nm which reduces interference caused by c...

Страница 4: ...hly with a lint free tissue HI 93703 70 before inserting the cuvette into the measurement cell The cuvette must be completely free of finger marks oil or dirt especially in the area where the light passes through i e approximately the bottom 2 cm 3 4 inch of the cuvette Place the cuvette into the holder and check that the notch on the cap is positioned securely into the groove OPERATIONAL GUIDE 1 ...

Страница 5: ...resence of floating debris and coarse sediments which may settle rapidly will give false readings The infrared light source of LP 2000 can effectively minimize errors due to colored substances found dissolved in solution This True Color effect is the most common interference found in commercial instruments operating in the visible light range Air bubbles and the effect of vibrations that disturb t...

Страница 6: ...nce more An intermittent ZERO will appear on the LCD CALIBRATION Using the HI 93703 0 bottle containing the ZERO FTU Standard or turbidity free water fill the measurement cuvette and secure the cap Note To minimize error caused by the cu vette it is recommended to use the same cuvette for calibration and mea surement Insert cuvette into the holder Press CAL An intermittent SIP indicates that the i...

Страница 7: ...dust or foreign matter that may be present Gently pour the liquid down the side of the vial to reduce air bubbles no mixing is required when Hanna standards are used Before inserting the vial into the meter wipe it clean with HI93703 70 or a soft lint free tissue Handle vials so that no fingerprints are left on the areas where light passes through bottom 2 cm STANDARD SUSPENSION Presently there ar...

Страница 8: ...EPA 1 10 FTU 30 mL HI 93703 50 Cuvette cleaning solution 230 mL ACCESSORIES WARRANTY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions This warranty is limited to repair or replacement free of charge Damage due to accident misuse tampering or lack of prescribed maintena...

Страница 9: ...942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 66 2619 0708 Fax 66 2619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765 7500 Fax 401 765 7575 For e mai...

Отзывы: