i) Socket for 12 Vdc power adapter

FUNCTIONAL DESCRIPTION

* 1 FTU = 1 NTU

Hanna Instruments uses the primary standard AMCO-AEPA-1 to
avoid all formazine-related problems. Formazine is a toxic, unstable
substance, which requires particular care. Calibration requiring formazine
standards must be preformed immediately after preparation of the
standards. Formazine standards may not be reused because of their
short life. 

LP 2000

 standards are extremely stable and have a shelf

life of up to six months. Uncontaminated standards may be reused.

LP 2000

 can be used with both AMCO-AEPA and formazine stan-

dards.
FTU is equal to the NTU. However, other known measurements
include: Jackson Turbidity Unit or JTU (based on the old method of
Jackson's candle), and Silica Unit in mg/L of SiO

2

. See the conversion

table below for these units:

JTU

FTU/NTU

SiO

(mg / L)

JTU

1

19

2.5

FTU/NTU

0.053

1

0.13

SiO

(mg / L)

0.4

7.5

1

MEASURE UNITS

LP 2000

 has been designed to perform measurements according to

the ISO 7027 International Standards.
The instrument functions by passing a beam of infrared light through
a vial containing the sample being measured.
The light source is a High Emission Infrared LED with a wavelength
peaking at 890 nm, which reduces interference caused by colored
samples to a minimum.
A sensor, positioned at 90° with respect to the direction of light,
detects the amount of light scattered by the undissolved particles
present in the sample. The microprocessor converts such readings into
FTU values.

IR LED

EMITTED LIGHT (890 nm)

CUVET

LIGHT

DETECTOR

90° SCATTERED

LIGHT

PRINCIPLE OF OPERATION

1) Cuvette holder
2) Dual level LCD (Liquid Crystal Display)
3) ON/OFF key, to switch the meter ON and OFF
4) CAL key, to enter calibration mode
5) READ/

, to perform measurements, and set the day and

month of the last calibration

6) DATE/

   

, to display the last calibration date, and select the

month or day of the last calibration

Содержание LP 2000

Страница 1: ...Instruction Manual LP 2000 Precision Bench Turbidity Meter w w w h a n n a i n s t c o m ...

Страница 2: ...LCD The meter is simple to use all operations can be carried out with only four keys and troubleshooting functions can be performed with error codes on the display An exclusive positive locking system guarantees that the cuvette is firmly placed in the cell The keypad is water resistant and can be wiped clean Calibration at 10 FTU can be easily performed using the supplied standard In addition LP ...

Страница 3: ...1 19 2 5 FTU NTU 0 053 1 0 13 SiO2 mg L 0 4 7 5 1 MEASURE UNITS LP 2000 has been designed to perform measurements according to the ISO 7027 International Standards The instrument functions by passing a beam of infrared light through a vial containing the sample being measured The light source is a High Emission Infrared LED with a wavelength peaking at 890 nm which reduces interference caused by c...

Страница 4: ...hly with a lint free tissue HI 93703 70 before inserting the cuvette into the measurement cell The cuvette must be completely free of finger marks oil or dirt especially in the area where the light passes through i e approximately the bottom 2 cm 3 4 inch of the cuvette Place the cuvette into the holder and check that the notch on the cap is positioned securely into the groove OPERATIONAL GUIDE 1 ...

Страница 5: ...resence of floating debris and coarse sediments which may settle rapidly will give false readings The infrared light source of LP 2000 can effectively minimize errors due to colored substances found dissolved in solution This True Color effect is the most common interference found in commercial instruments operating in the visible light range Air bubbles and the effect of vibrations that disturb t...

Страница 6: ...nce more An intermittent ZERO will appear on the LCD CALIBRATION Using the HI 93703 0 bottle containing the ZERO FTU Standard or turbidity free water fill the measurement cuvette and secure the cap Note To minimize error caused by the cu vette it is recommended to use the same cuvette for calibration and mea surement Insert cuvette into the holder Press CAL An intermittent SIP indicates that the i...

Страница 7: ...dust or foreign matter that may be present Gently pour the liquid down the side of the vial to reduce air bubbles no mixing is required when Hanna standards are used Before inserting the vial into the meter wipe it clean with HI93703 70 or a soft lint free tissue Handle vials so that no fingerprints are left on the areas where light passes through bottom 2 cm STANDARD SUSPENSION Presently there ar...

Страница 8: ...EPA 1 10 FTU 30 mL HI 93703 50 Cuvette cleaning solution 230 mL ACCESSORIES WARRANTY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions This warranty is limited to repair or replacement free of charge Damage due to accident misuse tampering or lack of prescribed maintena...

Страница 9: ...942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 66 2619 0708 Fax 66 2619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765 7500 Fax 401 765 7575 For e mai...

Отзывы: