background image

11

10

Calibration is recommended at least once a month. You can verify the
instrument calibration by comparing the display reading against the
supplied standard solution at 10 FTU. To check the date of the last
calibration, simply hold the DATE/    key down for a few seconds.

CALIBRATION PROCEDURE

Turn the meter on and wait for the display
to show "----".

Press the CAL key once. "CAL" will blink on the display for approx.
6 seconds, allowing for confirmation.

While "CAL" is still blinking, press CAL again (otherwise, it exits
this mode). "CL" will appear on the lower part of the display.

The date of calibration may now be edited
by simply pressing the DATE/    key. The
default parameter, which is the month
segment found on the left, will blink. To
scroll to the correct number, press the
READ/

s

 key. To toggle between month

and the day display segments, press the
DATE/    key. The day segment can also
be edited with the READ/

s

 key.

To confirm the displayed values and to proceed, press CAL once
more. An intermittent "ZERO" will appear on the LCD.

CALIBRATION

Using the 

HI 93703-0 

bottle containing

the ZERO FTU Standard (or turbidity-free
water), fill the measurement cuvette and
secure the cap.

Note

: To minimize error caused by the cu-

vette, it is recommended to use the
same cuvette for calibration and mea-
surement.

Insert cuvette into the holder.

• Press CAL. An intermittent “SIP” indicates that the instrument is

being calibrated.

• After approximately 1 minute, the instru-

ment will ask for the 

HI 93703-10

standard solution @ 10 FTU by display-
ing “10.0”.

Refill the cuvette with the 

HI 93703-10

AMCO-AEPA-1 standard at 10 FTU and
place it into the measurement cell

After the second calibration point (10
FTU) has been accepted, the meter will
display “500”, asking for the 500 FTU
solution to be placed in the cuvet holder.

Note:

 At this point the user can exit the calibration mode and save

the two-point calibration by pressing READ.

To perform a three-point calibration, place the 500 FTU standard
solution in the cuvet holder.

Press CAL: “SIP” and “CL” will start
blinking.

After approximately 30 seconds, the dis-
play will show "----".

Now the meter is calibrated and ready for use.

Содержание LP 2000

Страница 1: ...Instruction Manual LP 2000 Precision Bench Turbidity Meter w w w h a n n a i n s t c o m ...

Страница 2: ...LCD The meter is simple to use all operations can be carried out with only four keys and troubleshooting functions can be performed with error codes on the display An exclusive positive locking system guarantees that the cuvette is firmly placed in the cell The keypad is water resistant and can be wiped clean Calibration at 10 FTU can be easily performed using the supplied standard In addition LP ...

Страница 3: ...1 19 2 5 FTU NTU 0 053 1 0 13 SiO2 mg L 0 4 7 5 1 MEASURE UNITS LP 2000 has been designed to perform measurements according to the ISO 7027 International Standards The instrument functions by passing a beam of infrared light through a vial containing the sample being measured The light source is a High Emission Infrared LED with a wavelength peaking at 890 nm which reduces interference caused by c...

Страница 4: ...hly with a lint free tissue HI 93703 70 before inserting the cuvette into the measurement cell The cuvette must be completely free of finger marks oil or dirt especially in the area where the light passes through i e approximately the bottom 2 cm 3 4 inch of the cuvette Place the cuvette into the holder and check that the notch on the cap is positioned securely into the groove OPERATIONAL GUIDE 1 ...

Страница 5: ...resence of floating debris and coarse sediments which may settle rapidly will give false readings The infrared light source of LP 2000 can effectively minimize errors due to colored substances found dissolved in solution This True Color effect is the most common interference found in commercial instruments operating in the visible light range Air bubbles and the effect of vibrations that disturb t...

Страница 6: ...nce more An intermittent ZERO will appear on the LCD CALIBRATION Using the HI 93703 0 bottle containing the ZERO FTU Standard or turbidity free water fill the measurement cuvette and secure the cap Note To minimize error caused by the cu vette it is recommended to use the same cuvette for calibration and mea surement Insert cuvette into the holder Press CAL An intermittent SIP indicates that the i...

Страница 7: ...dust or foreign matter that may be present Gently pour the liquid down the side of the vial to reduce air bubbles no mixing is required when Hanna standards are used Before inserting the vial into the meter wipe it clean with HI93703 70 or a soft lint free tissue Handle vials so that no fingerprints are left on the areas where light passes through bottom 2 cm STANDARD SUSPENSION Presently there ar...

Страница 8: ...EPA 1 10 FTU 30 mL HI 93703 50 Cuvette cleaning solution 230 mL ACCESSORIES WARRANTY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions This warranty is limited to repair or replacement free of charge Damage due to accident misuse tampering or lack of prescribed maintena...

Страница 9: ...942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 66 2619 0708 Fax 66 2619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765 7500 Fax 401 765 7575 For e mai...

Отзывы: