•  Mechanical cleaning of the membrane with abrasives is not recommended.

•  Clean the exterior of the probe by wiping with a aqueous-soapy mixture and 

gently rubbing persistent spots off. Rinse with clean water. 

•  Replace the cap and electrolyte if coating persists or membrane damage is evident.

Membrane Cap & Electrolyte Replacement

The membrane cap and electrolyte are designed to provide trouble-free operation 

for up to eight weeks. Replacements are required when: 

•  The membrane cap is physically damaged 

•  Probe response is slower than usual, especially after many months of use 

•  DO probe calibration or readings exhibit greater than normal drift, especially 

after many months of use 

•  The membrane cap remains fouled after cleaning procedures were implemented.

Procedure

1.  Remove probe from installation.

2.  Unscrew and remove electrolyte screw located on the side of the probe body. 

3.  While holding the probe assembly vertical, unscrew the membrane cap 

assembly from the probe body. Discard the old membrane cap assembly. To 

remove old electrolyte, shake probe down. 

4.  Flush the probe body and electrolyte reservoir with tap water. Dry assembly 

with a lint-free cloth or tissue. 

5.  Inspect O-rings for nicks or wear. Discard and replace damaged ones.

6.  If tarnished or stained, gently clean the cathode with a lint-free cloth.

7.  Follow section 6 of this manual for filling instructions.

Long Term Storage

Discard any electrolyte solution from the reservoir and flush probe body and 

reservoir with water. Store the probe with the protective cap on.

11.  ACCESSORIES

Ordering Information

Product Description

DO Solutions

HI7040L

Zero oxygen solution set, 500 mL + 12 g

HI7042S

Refilling electrolyte solution, 30 mL

Other Accessories

HI60501

PVC immersion electrode holder

HI60501-0

O-rings for 

HI60501

 electrode holder

HI605011

PVC mounting flange for 

HI60501

 electrode holder

HI60503

PVDF immersion electrode holder

HI60542

In-line electrode holder, direct pipe installation

HI740155P

Electrode refilling capillary pipette (20 pcs.)

HI76510-05

Probe extension cable, 5 m (16’5”) long

HI76510-10

Probe extension cable, 10 m (32’9”) long

HI76510-15

Probe extension cable, 15 m (49’2”) long

HI76510-25

Probe extension cable, 25 m (82’) long

HI76510-50

Probe extension cable, 50 m (164’) long

BL120-400

Flow-cell probe adapter kit

BL120-401

Flow-cell valve

BL120-402

Flow-cell tubing,10 m (32’8”) long

BL120-410

Flow cell

BL120-450

Flow-cell kit for Ø 50 mm (2”) pipe

BL120-463

Flow-cell kit for Ø 63 mm (2.5”) pipe

BL120-475

Flow-cell kit for Ø 75 mm (3”) pipe

BL120-500

Probe fitting kit

BL120-550

Probe saddle for Ø 50 mm (2”) pipe, 1¼” thread

BL120-563

Probe saddle for Ø 63 mm (2.5”) pipe, 1¼” thread

BL120-575

Probe saddle for Ø 75 mm (3”) pipe, 1¼” thread

For additional accessories information contact your local sales office.

Содержание HI7640-18 Series

Страница 1: ...INSTRUCTION MANUAL HI7640 18 Series Galvanic Dissolved Oxygen Industrial Smart Probe dedicated to HI510 Process Controller HI7640 1805 SN XXXXXXXXXXXX 0 to 50 ppm mg L T 5 to 50 C p max 3 bar...

Страница 2: ...a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument a...

Страница 3: ...Safety Measures 4 3 General Description 4 4 Main Features 5 5 Specifications 5 6 Preparing for Use 6 7 Installation 6 8 Wiring the Probe to the Controller 8 9 Calibration 9 10 Maintenance 9 11 Access...

Страница 4: ...tructions in this manual 3 GENERAL DESCRIPTION The HI7640 18 is a galvanic style oxygen probe series designed for use with the Hanna Instruments HI510 Universal Process Controller The probe is suited...

Страница 5: ...5 0 to 50 0 C 23 0 to 122 0 F Temperature accuracy 0 5 C 1 0 F Temperature compensation 5 0 to 50 0 C 23 0 to 122 0 F Body PVDF Sensor type Galvanic cell Cathode Silver Ag Anode Zinc Zn Water movement...

Страница 6: ...he fill hole aligned upright slowly insert the sensing tip into the membrane cap so excess electrolyte and any remaining trapped air escapes through the fill hole 9 Screw the membrane cap onto sensor...

Страница 7: ...ugh the pipe cap opening 3 3 Screw the subassembly onto the PCV pipe 5 4 Tighten the cable locking system 2 to secure the cabling When removing the probe from tank fish tank water treatment tank or sa...

Страница 8: ...thread 8 WIRING THE PROBE TO THE CONTROLLER 1 Run the probe cable through the conduit opening 2 Connect the probe leads to the removable terminal connector marked PROBE Follow the lead markings positi...

Страница 9: ...hen displayed followed by Next 4 Submerge the probe into HI7040 Zero oxygen solution and stir gently for 2 3 minutes Wait for the temperature and probe values to become stable The meter will automatic...

Страница 10: ...clean the cathode with a lint free cloth 7 Follow section 6 of this manual for filling instructions Long Term Storage Discard any electrolyte solution from the reservoir and flush probe body and reser...

Страница 11: ...used Any variation introduced by the user to the supplied equipment may degrade its performance Do not use or store the product in hazardous environments WARRANTY The probes are warranted for six mon...

Страница 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN7640 18 Printed in ROMANIA...

Отзывы: