Installation 8
8. INSTALLATION
General considerations
• Probes are easily installed using the
3/4
” NPT external thread.
• Hand tighten the probe in position. Then, depending on the process, tighten one or two turns with a wrench
to secure in place. Do not exceed 10 N
•
m (7.3 lb-ft) torque specification for the probe sensor.
• Provisions must be made for removing the probe from the process.
• Consider probe accessibility for maintenance when selecting placement.
Typical installation examples with mounting accessories
Accessories are sold separately.
In-line mounting and flow-cell installation require that the saddle and flow cell are completely filled with water.
In-line mounting with probe saddle
5
4
2
3
1
1 Pipe
2 Sensor tip
3 Saddle
4 Probe fitting kit
5 Probe
Note
: To prevent air entrapment inside the electrode the probe can be rotated and horizontal mounting
is recommended.
Mounting accessories
Pipe size
Saddle code
HI7630-28
zz
HI7630-48
zz
Ø 50 mm (2”)
BL120-550
Ø 63 mm (2 ½”)
BL120-563
Ø 75 mm (3”)
BL120-575