Wiring the Probe to the Controller

9.  WIRING THE PROBE TO THE CONTROLLER

1.  With the controller disconnected from power, run the probe cable through the conduit opening.
2.  Connect the probe leads to the removable terminal connector marked PROBE. Follow the lead markings 

(+) positive / (–) negative to ensure correct wiring position for the output leads.

3.  Carefully put the wired terminal connector into place on the board.
4.  Position excess cable through the cable gland before tightening the nut.
5.  Remove the ground screw and hardware located below the PROBE connector and attach ground lead ( 

).

COMM

- B A +

IN1 IN2  A01 A02  A03 A04 

PROBE

- B A +

PROBE

- B A +

A +

B

Probe cabling color code

Marking

Attached cable

Patch cable *

Functionality

GREEN

BLACK

0V

B

WHITE

WHITE

RS485 D –

A

YELLOW

BLUE

RS485 D+

+

BROWN

RED

5V

g

GREEN-YELLOW 

Protective ground connection

See Accessories section for cable ordering codes.

Note

: Ensure wiring regulations are correctly followed when controller unit is part of a larger industrial installation.

Содержание HI7630-28

Страница 1: ...4805 0 to 30 mS cm T 0 to 50 C p max 6 bar SN XXXXXXXXXXXX HI7630 2800 Conductivity and Temperature Industrial Smart Probes Please scan the QR code or follow the link to download the HI510 controller...

Страница 2: ...intenance 14 13 Accessories 15 14 Abbreviations 16 Certification 16 Recommendations for Users 16 Warranty 16 Dear Customer Thank you for choosing a Hanna Instruments product Please read this instructi...

Страница 3: ...ons detailed in this manual 3 GENERAL DESCRIPTION INTENDED USE HI7630 28 and HI7630 48 are Conductivity and Temperature probes designed to be paired with the HI510 Universal Process Controller The pro...

Страница 4: ...conductivity probes can be installed directly in line immersed in a tank or in a flow cell and are suited for continuous measurement of conductivity and associated parameters required in applications...

Страница 5: ...h standard solution Process Single point Temperature Range 0 0 to 50 0 C 32 0 to 122 0 F Accuracy 0 5 C 1 0 F Calibration Single point offset controller setting Temperature compensation mode Linear Na...

Страница 6: ...ion Salinity Single point with standard solution Resistivity Single point Process Single point Temperature Range 0 0 to 100 0 C 32 0 to 212 0 F Accuracy 0 5 C 1 0 F Calibration Single point offset con...

Страница 7: ...189 mm 7 4 Four Ring Probes HI7630 48zz 28 mm 1 1 72 mm 2 8 50 mm 2 12 mm 0 5 3 4 NPT 216 mm 8 5 HI7630 4800 216 mm 8 5 Align the pins and key then push the plug into the socket Rotate the collar to...

Страница 8: ...essibility for maintenance when selecting placement Typical installation examples with mounting accessories Accessories are sold separately In line mounting and flow cell installation require that the...

Страница 9: ...PT 2 Vent hole 3 O ring 4 Probe Tank immersion with submersible electrode holder I II III 1 2 4 3 5 6 7 9 8 1 Probe cable 2 Cable gland 3 Electrode holder cap 4 Adjustable immersion level 5 Electrode...

Страница 10: ...3 and cable gland on top 2 The cable is shielded inside the holder to prevent any damage to the insulation F Tighten the cable gland 2 step III and slide the cap 3 back onto the electrode holder 5 Fl...

Страница 11: ...n the board 4 Position excess cable through the cable gland before tightening the nut 5 Remove the ground screw and hardware located below the PROBE connector and attach ground lead g COMM B A IN1 IN2...

Страница 12: ...oss contamination when a two point calibration is required use two beakers one for rinsing the probe and the other for calibration Go to Probe Settings in Channel Setup and set Meas Mode to EC Procedu...

Страница 13: ...ed in conductivity EC Meas Mode The user can subsequently change Meas Mode to RES to measure in resistivity units M cm Ultrapure water self ionizes into H and OH ions and has a conductivity of 0 055 S...

Страница 14: ...ccurate readings Remove and inspect the probe during scheduled service intervals Dry clean the sensor with a soft bristle brush to loosen any debris For a more thorough cleaning Use a cloth and warm w...

Страница 15: ...1 2 PVC protective end cap inside height 68 mm 2 6 HI60503 PVDF immersion electrode holder HI60503 2 PVDF protective end cap inside height 68 mm 2 6 HI605011 PVC mounting flange HI60542 In line electr...

Страница 16: ...oduced by the user to the supplied equipment may degrade its performance Do not use or store the product in hazardous environments WARRANTY The probes are warrantied for six months against defects in...

Отзывы: