14. ERROR MESSAGES
The instrument shows clear warning messages (bottom of the screen) when erroneous conditions appear
or when measured values are outside the expected range.
The information below provides an explanation of the errors and warnings, and recommended action to be taken.
Displayed Message
Explanation & Recommended Action
“Temperature under/over range” Temperature outside specified range.
Replace probe if necessary.
“Under/over compensation range” During conductivity calibration, the temperature is under/over the
conductivity calibration solution compensation limit.
”EC under/over range”
Outside range in the corresponding scale.
Ensure the vent holes are completely submerged and the sample is within
specified range.
”Broken electrode”
The meter fails to calibrate or gives faulty readings.
Replace the probe.
”EC out of calibration range”
Displayed when the measured value is outside calibration range.
”Factory calibration expired”
Contact the Hanna technical support for the periodic factory calibration.
15. ACCESSORIES
CONDUCTIVITY SOLUTIONS
HI7030M or HI7030L
12880 μS/cm standard solution, 250 or 500 mL
HI7031M or HI7031L
1413 μS/cm standard solution, 230 or 500 mL
HI7033M or HI7033L
84 μS/cm standard solution, 230 or 500 mL
HI7034M or HI7034L
80000 μS/cm standard solution, 250 or 500 mL
HI7035M or HI7035L
111800 μS/cm standard solution, 230 or 500 mL
HI7037M or HI7037L
100 % NaCl seawater salinity standard solution, 250 or 500 mL
HI7039M or HI7039L
5000 μS/cm standard solution, 250 or 500 mL
OTHER ACCESSORIES
HI740036P
100 mL beaker (10 pcs.)
HI740037P
20 mL beaker (10 pcs.)
HI764060
Electrode holder
HI900946
115 Vac to24 Vdc power adapter, US plug
HI900947
230 Vac to 24 Vdc power adapter, European plug
HI920016
USB type A to C cable