Configured parameter icon is displayed i.e.
EC
Res
TDS
Sal
.
`
Autohold
is displayed blinking until the measurement is stable.
`
When stable, the indicator stops blinking and measurement freezes at current value.
`
Tap parameter icon to return to direct measurement.
9.3. TEMPERATURE
Options: Temperature Source, Temperature Unit, Temperature Compensation, Manual, Reference
Temperature, Temperature Coefficient, User Temperature Calibration
Temperature Source: Automatic, Manual
Users can select between physical temperature input source (Automatic) and entering sample temperature
value manually (Manual).
•
Automatic
: built‑in temperature sensor adjusts the measured conductivity to a reference temperature.
”ATC“is displayed indicating temperature compensation status.
•
Manual
: sample temperature is entered manually and displayed “MTC” indicates temperature compensation
status.
Measurement & Probe Setup Menu
40