12. LOGGING
T
ap
0%
to check available storage.
Three logging types are available: Automatic, Manual, and Autohold.
Automatic logging
• Readings are logged ( ) at predefined period intervals.
Sampling interval options range from 1 second to 180 minutes.
• Records are continuously added to it until the session stops.
• For each automatic logging session, a new lot is created.
• A complete set of GLP information including date, time, mode selection, temperature reading, and calibration
information is stored with each log. User and log information e.g. company, address, sample details may
also be included on the .CSV file.
Manual logging
• Readings are logged each time
M
is tapped.
• All manual readings are stored in a single log (i.e. records made on different days share the same log).
Autohold logging
• Readings are logged each time
D
is tapped and configured stability criteria is reached.
• All Autohold readings are stored in a single log (i.e. records made on different days are logged in the
same lot).
Note:
Stability criteria can be set to Fast, Medium, or Accurate for all logging types.
Logging