38
ERROR CODES
HI98713
has a powerful diagnostic system. The common errors are detected and reported for easy
diagnostic and maintenance.
ERROR
DESCRIPTION
ACTION
Err1 – Err3;
Err6; Err7; Err8
Critical errors.
The instrument beeps and shuts down.
Call Hanna Instruments service
Err4
The instrument beeps shortly twice and
shuts down after 10 seconds.
Press simultaneously UP and
DOWN to reset the EEPROM
contents.
CAP
The lid is not closed.
Close the lid.
If the error persists, return the
instrument.
no L
Lamp broken or no light.
Replace the lamp. Check the
optical system for obstructions.
L Lo
Not enough light.
Check the optical system for
obstructions.
-LO-
The standard used for current calibra‑
tion point is too low.
Check the standard and use
the correct one.
-HI-
The standard used for current calibra‑
tion point is too high.
Check the standard and use
the correct one.
Battery tag blinking
The remaining battery life is too low. Replace batteries.
bAtt
The batteries are too discharged for
correct measurements.
Replace batteries.
PC INTERFACE
To fully use the instrument tag identification system function, the measured data has to be downloaded
to a computer. The instrument can use RS232 or USB connection to communicate with the PC.
When using the RS232 protocol, simply connect a
HI920011
serial cable between the instrument
and the computer.
To use the USB protocol, simply connect a regular USB cable between instrument and PC.
In both cases, the PC must run the
HI92000
application for successful data transfer. To allow our users
access to the latest version of Hanna Instruments PC compatible software, we made the products available
for download at
http://software.hannainst.com
. Select the product code and click
Download
Now
.