26
LOGGING
LOGGING
The log function is active after a valid measurement
is obtained (no errors).
• To log a value, press LOG/CFM when the mea‑
surement result is displayed.
The instrument asks to READ TAG for identification
of the sampling location. The location for the new
record is also displayed on the secondary LCD.
• To read the ID code for the sampling location
identification, simply touch the iButton
®
tag with
the matching connector, located on the back of
the instrument (see Connectors Description, page
9). Alternatively, press again LOG/CFM to store
the record without the tag ID code.
• If the tag is successfully read, the instrument will
beep once, displaying the unique hexadecimal
code of the tag, and store the data.
After data is stored, the instrument returns to
measurement mode.
Notes
: • If the tag is not read within 20 seconds,
the logging procedure is canceled.
• A measurement can be stored only once.
Also an over range value can be stored.
• If less than ten free records are available, the
“LOG” tag will blink while storing data.
• If the log memory is full, the “LoG FULL” message
will appear for a few seconds on the LCD and
the instrument will return to measurement mode
without storing the new record.
To store a new record, delete one or more records.
HI98713
has a log space for up to 200 records. With each measurement, the date, time and tag ID
is stored. In this way, each record is fully characterized and can be easily analyzed when downloading
data on the PC application (
HI92000
).
or