27
VIEW LOGGED DATA
The stored records can be viewed at any moment
by pressing RCL. To return to normal measurement
mode, press RCL again .
LOG SEARCHING
The log records are stored in chronological order.
The first displayed record is the last stored one.
• Press UP or DOWN keys to scroll the log memory
record by record. By keeping pressed the UP or
DOWN keys, the scrolling speed will increase.
The scrolling of the log is possible from any
panel of the record, except “Delete last log”
and “Delete all logs” panels.
• When scrolling the log, the number of the record
is displayed for one second on the secondary
LCD, together with “TAG” if the identification
of the sampling location was made.
When the end of the log is reached, an error beep
is played.
RECORD VIEWING
Each record contains more information than the
measured value. The additional information is
grouped in several panels.
Press READ to scroll through the record panels.
The record panels are displayed one by one in a
circular way.
Each record contains the following panels:
• The record value (turbidity value) and record
number.
Note
: If the logged sample value is an over range
reading, the maximum value (1000) will be
displayed blinking.
• The hexadecimal string of the tag for the
sampling location ID.
Note
: If the ID data are missing, dashes are
displayed instead.
or