INITIAL PREPARATION

Each meter is supplied complete with batteries. Remove the back
cover, unwrap the batteries and install them while paying attention
to the polarity. Alternatively, connect the supplied 12 Vdc adapter to
the power adapter plug.
To prepare the instrument for use, connect the pH or ORP (

HI 98240

only) electrode to the connector located on the top of the instrument.

To switch the meter on, press the ON/OFF key.
The batteries charge status or "LINE" message
(if external power adapter is connected) will be
displayed on the LCD for a few seconds.
The meter is now ready to operate.

To maximize battery life, the meter is automatically switched off after a
user selectable  period of non-use (default is 5 minutes). If in logging
mode, after the period of non-use, the meter will continue to monitor
pH or mV (

HI 98240 

only) and temperature periodically at the end of

every logging interval. Only the "LOG" indication will be visible on
LCD. While storing data in memory, during the sleep mode, the
reading will appear briefly on the LCD.
Before proceeding with pH measurements follow the calibration proce-
dure.

Note:

When the use of an alternate function (FNC, CFM and CAL) is
requested, press and hold the ALT key first and then the
second listed key.

Note:

 To prevent damage to the electrode, remove the pH electrode

from the sample before turning the meter off.

If the meter is OFF, detach the electrode from the meter before
immersing the electrode in the storage solution.

ON/OFF

SPECIFICATIONS

Range

-4.00 to 19.99 pH

±400.0 mV / ±2000 mV autoranging (

HI 98240

 only)

-10.0 to 120.0°C / 14.0 to 248.0°F

Resolution

0.01 pH

0.1 mV (±400 mV) / 1 mV (outside) (

HI 98240

 only)

0.1°C / 0.1°F

Accuracy (@20°C/68°F)

±0.01 pH

±0.5 mV (±400 mV) / ±2 mV (outside) (

HI 98240

 only)

±0.5 °C (0 to 70°C) / ±1°C (outside)

±1°F (32 to 158°F) / ±1.8°F (outside)

Typical EMC Deviation

±0.03 pH

±1 mV (±400 mV) / ±2 mV (outside) (

HI 98240

 only)

±0.8°C / ±1.4°F

pH Calibration

Automatic, 1 or 2 point with 5 memorized buffer values

(pH 4.01, 6.86, 7.01, 9.18 and 10.01)

mV Calibration

Automatic, 2 point at 0 and 350 mV or

(

HI 98240

 only)

 3 point at 0, 350 and 1900 mV

Temperature Compensation

Automatic, -10 to 120°C (14 to 248°F)

pH Electrode

HI 1618D

 smart electrode

with built-in temperature, DIN connector and 1 m (3.3') cable (included)

ORP Electrode

Samrt ORP electrode with

(

HI 98240

 only)

DIN connector and 1 m (3.3') cable (optional)

Input Impedance

10

12 

Ohm

PC Connection

HI 9200

 IR interface

Printer

Low power impact type-belt, 14 characters per line;

38 mm plain paper (

HI 710034

)

Printing/Logging Interval

Selectable at 1, 2, 5, 10, 15, 30, 60, 120 or 180 minutes

Power supply

4 x 1.5V AA batteries / approx. 350 of use

(without printing and backlight)

or 12 Vdc input

Environment

0 to 50°C (32 to 122°F); RH max 95%

Dimensions

220 x 87 x 75 mm (8.7 x 3.4 x 2.9")

Weight

550 g (1.2 lb.)

Содержание HI 98230

Страница 1: ...Instruction Manual HI 98230 HI 98240 Printing and Logging pH Meters with GLP and Smart Electrodes w w w h a n n a i n s t c o m...

Страница 2: ...meters are also equipped with a stability indicator and backlight feature for comfortable reading even in dark environments The GLP features provide a guarantee of data consistency Measurements can b...

Страница 3: ...the LOBAT indication is displayed to warn the user of this condition However the meter continues to measure correctly even when the low battery indication is displayed When the batteries become too we...

Страница 4: ...lectrode remove the pH electrode from the sample before turning the meter off If the meter is OFF detach the electrode from the meter before immersing the electrode in the storage solution ON OFF SPEC...

Страница 5: ...iption of the specific setup items their valid values and the factory settings default Code Valid values Default 00 Lot data printing scrolling 00 to 16 00 01 Print lots data summary 02 Printer enable...

Страница 6: ...sample is reached press and hold the ALT and PAPER keys until the printer stops During printout the LCD will display the sample number being Code Valid values Default 10 Show pH GLP data 11 Show ORP G...

Страница 7: ...age GLP data printing will scroll across LCD The meter will then print a complete set of the following GLP data a Instrument identification ID code printed at that moment If printout is stopped the LC...

Страница 8: ...ferent samples it is recommended to rinse the electrode thoroughly with deionized water or if not available tap water first and then with some of the next sample to condition the electrode before imme...

Страница 9: ...ORP electrode to the meter and press ON OFF to power on the instrument The meter automatically sets the mV range If the ORP electrode is not connected the LCD will display the no probe message and th...

Страница 10: ...ng will be printed beside the date Note A dashed line indicates that the calibration has never been performed 6 If password is correct or not set the meter will display 7 01pH on the up per LCD and b1...

Страница 11: ...when LCD is displaying mV 3 User is prompted to enter password if it has been set to a value different from 0000 otherwise skip to step 6 4 Enter the password with the arrow keys 5 Press ALT and CFM t...

Страница 12: ...ly enabled in order to return to logging with printing mode LOG ON DEMAND In measuring or Timed log mode press LOG to store the current reading The LCD will display Two different modes to print log da...

Страница 13: ...the meter on again CALIBRATION ALARM TIME OUT The calibration alarm time out is available only for pH calibration It is possible to set through setup code 13 the number of days before the next requir...

Страница 14: ...t is possible to make any elaboration available with software e g graphics statistic analysis HI 92000 offers a variety of features and is provided with an on line help to support the user throughout...

Страница 15: ...lot was the 16th the new logging period restarts from lot 01 overwriting previously logged data When Timed logging memory is full the meter overwrites the oldest lot data progressively reducing the o...

Страница 16: ...I 98240 can also be used with other adapters In this case remember to check the correct polarity of your adapter before connecting it to the meter BATTERY REPLACEMENT TEMPERATURE RESISTANCE CORRELATIO...

Страница 17: ...own could be due to Clogged Dirty Junction refer to the Cleaning Procedure above Loss of shielding due to low electrolyte level in refillable electrodes only refill with fresh HI 7071 for single junct...

Страница 18: ...7 01 buffer solution 500 mL FDA bottle HI 8009L pH 9 18 buffer solution 500 mL FDA bottle HI 7009L pH 9 18 buffer solution 500 mL bottle HI 7010L pH 10 01 buffer solution 500 mL bottle HI 8010L pH 10...