15

14

Connect the pH electrode with the built-in
temperature sensor to the meter and press
ON/OFF to power on the instrument.
The meter automatically checks that the stored calibration data
correspond to the connected electrode. The pH range will be auto-
matically set.
If the pH electrode is not connected, the LCD will display the "no
probe" message and then dashed lines in place of the readings.
For greatest accuracy, it is recommended to set
the Calibration alarm time-out to the value
appropriate to your specific use and calibrate
the meter as soon as the "DATE" warning
symbol blinks on LCD (see page 25).
To take pH measurements, simply submerge
the bottom 4 cm (1½") of the electrode in
the solution to be tested, stir gently and
allow for the reading to stabilize. The tem-
perature is displayed in the lower LCD without
the decimal digit.
The pH reading is automatically temperature compensated (ATC).
If the pH reading blinks it means that the electrode is "dead" and
readings are not reliable.
By continuously pressing the RANGE key the follow-
ing information will be displayed on the upper LCD:
• mV reading (

HI 98240

 only)

The mV scale is autoranging, when the
reading is outside ±400 mV the deci-
mal point automatically disappears.

• Temperature reading

 If  the temperature reading is out of

range, a dashed line will be displayed.

• Date
• Time
Pressing RANGE again, the meter returns to pH reading.
If measurements are taken successively in different samples, it is
recommended to rinse the electrode thoroughly with deionized water
or, if not available, tap water first and then with some of the next
sample to condition the electrode before immersing it in the sample.

TAKING pH MEASUREMENTS

b - last pH calibration date
c - last pH calibration time
d - pH electrode offset
e - pH electrode slope
f

-  1

st

 point pH calibration buffer

g -  2

nd

 point pH calibration buffer

(if present)

h - last mV calibration date

(

HI98240

 only)

i

- last mV calibration time

(

HI98240

 only)

l

-  1

st

 point mV calibration (

HI98240

 only)

m -  2

nd

 point mV calibration (

HI98240

 only)

n -  3

rd

 point mV calibration (if present; 

HI98240

 only).

Note:

If the pH is followed by a set of * or -, it means  that the
pH electrode is old or dead respectively.
A dashed line indicates that the calibration has never been
performed.

If the calibration date is followed by the symbol "Dt", it means
that the calibration alarm time-out period has been exceeded.

CODE 13 - Setting the Identification (ID) Code

When using several identical meters it may be useful to uniquely
identify them by assigning an ID code to each meter.
• Select code 13. The message "Instrument ID code" will scroll across

LCD.

• Enter a 4-digit value using the arrow keys.
• Press ALT and CFM to confirm the value.

TESTING BATTERY LEVEL

• Select code 41. The message "Battery test" will scroll across LCD.
• If the meter is connected to an external power adapter, the LCD

will display "LINE", otherwise it will display bAtt on the upper
display, and the remaining percentage
of battery charge (100% means fully
charged battery and 0% corresponds to
the minimum battery voltage that al-
lows the meter to operate).

Содержание HI 98230

Страница 1: ...Instruction Manual HI 98230 HI 98240 Printing and Logging pH Meters with GLP and Smart Electrodes w w w h a n n a i n s t c o m...

Страница 2: ...meters are also equipped with a stability indicator and backlight feature for comfortable reading even in dark environments The GLP features provide a guarantee of data consistency Measurements can b...

Страница 3: ...the LOBAT indication is displayed to warn the user of this condition However the meter continues to measure correctly even when the low battery indication is displayed When the batteries become too we...

Страница 4: ...lectrode remove the pH electrode from the sample before turning the meter off If the meter is OFF detach the electrode from the meter before immersing the electrode in the storage solution ON OFF SPEC...

Страница 5: ...iption of the specific setup items their valid values and the factory settings default Code Valid values Default 00 Lot data printing scrolling 00 to 16 00 01 Print lots data summary 02 Printer enable...

Страница 6: ...sample is reached press and hold the ALT and PAPER keys until the printer stops During printout the LCD will display the sample number being Code Valid values Default 10 Show pH GLP data 11 Show ORP G...

Страница 7: ...age GLP data printing will scroll across LCD The meter will then print a complete set of the following GLP data a Instrument identification ID code printed at that moment If printout is stopped the LC...

Страница 8: ...ferent samples it is recommended to rinse the electrode thoroughly with deionized water or if not available tap water first and then with some of the next sample to condition the electrode before imme...

Страница 9: ...ORP electrode to the meter and press ON OFF to power on the instrument The meter automatically sets the mV range If the ORP electrode is not connected the LCD will display the no probe message and th...

Страница 10: ...ng will be printed beside the date Note A dashed line indicates that the calibration has never been performed 6 If password is correct or not set the meter will display 7 01pH on the up per LCD and b1...

Страница 11: ...when LCD is displaying mV 3 User is prompted to enter password if it has been set to a value different from 0000 otherwise skip to step 6 4 Enter the password with the arrow keys 5 Press ALT and CFM t...

Страница 12: ...ly enabled in order to return to logging with printing mode LOG ON DEMAND In measuring or Timed log mode press LOG to store the current reading The LCD will display Two different modes to print log da...

Страница 13: ...the meter on again CALIBRATION ALARM TIME OUT The calibration alarm time out is available only for pH calibration It is possible to set through setup code 13 the number of days before the next requir...

Страница 14: ...t is possible to make any elaboration available with software e g graphics statistic analysis HI 92000 offers a variety of features and is provided with an on line help to support the user throughout...

Страница 15: ...lot was the 16th the new logging period restarts from lot 01 overwriting previously logged data When Timed logging memory is full the meter overwrites the oldest lot data progressively reducing the o...

Страница 16: ...I 98240 can also be used with other adapters In this case remember to check the correct polarity of your adapter before connecting it to the meter BATTERY REPLACEMENT TEMPERATURE RESISTANCE CORRELATIO...

Страница 17: ...own could be due to Clogged Dirty Junction refer to the Cleaning Procedure above Loss of shielding due to low electrolyte level in refillable electrodes only refill with fresh HI 7071 for single junct...

Страница 18: ...7 01 buffer solution 500 mL FDA bottle HI 8009L pH 9 18 buffer solution 500 mL FDA bottle HI 7009L pH 9 18 buffer solution 500 mL bottle HI 7010L pH 10 01 buffer solution 500 mL bottle HI 8010L pH 10...

Страница 19: ...10005 Voltage adapter 115 Vac 12 Vdc HI 710006 Voltage adapter 230 Vac 12 Vdc HI 710031 Rugged carrying case HI 710034 Paper roll 10 pcs HI 710035 Ink cartridge HI 76405 Electrode holder HI 8427 pH mV...

Страница 20: ...s in workmanship and materials when used for their intended purpose and maintained according to instructions The electrodes and the probes are warranted for a period of six months This warranty is lim...

Страница 21: ...42 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 0...

Отзывы: