background image

27

26

Windows

®

 is registered Trademark of "Microsoft Co."

Excel

©

 Copyright "Microsoft Co."

Lotus 1-2-3

©

 Copyright "Lotus Co."

HI 98230

 and 

HI 98240

 feature a infrared transmitting circuitry.

Ensure there is not any logging process active.
Press RANGE to set the meter to time or date mode and simply place
your data-logger on a 

HI 9200

 infrared transmitter (ensuring that

the two infrared LEDs are placed on top of each other) and the
memory content can then be downloaded to your PC through a serial
port. Just ensure that baud rate on instrument (setup code 50) and
on PC downloading program are set to the same value.
During the data transfer the instrument
displays the message "r 232".

To stop communication, press RANGE to display the reading (pH or
mV or temperature) or take the meter out of the transmitter when it
is not displaying "r232".

Using the 

HI 9200

 infrared transmitter, all recorded data can be fed

to your PC for easy reproduction, storage or elaboration without the
interference of cables or cords between the meter and the transmitter.
Data transmission from the instrument to the PC is now much easier
with new 

HI 92000 

Windows

®

 compatible software by HANNA.

HI 92000 

allows to use the powerful capabilities of most spread sheet

programs (e.g. Excel

©

, Lotus 1-2-3

©

). Simply open the file down-

loaded by 

HI 92000 

from the spread sheet program and then it is

possible to make any elaboration available with software (e.g.
graphics, statistic analysis).

HI 92000 

offers a variety of features and is provided with an on-line-

help to support the user throughout any situation.
To install 

HI 92000 

you need a 3.5" drive and a few minutes to

follow the instructions conveniently printed on the disk label.

DATA TRANSFER TO PC

The meter checks if the time-out time has
expired every time it is turned on. In the
case the time has run out, the message "Cal
date" scrolls across the LCD. The "DATE" sym-
bol will blink as a reminder.
Alarm time-out is also signaled in the GLP
data printout through the "Dt" symbol after
the calibration date.

GLP AND RS232

All the GLP data can be retrieved from a PC through the RS232
communication feature. The calibration data are transferred to the
computer along with measurement data for analysis and verification
(see "Data transfer to PC" paragraph).

HI 98230

 and 

HI 98240

 are factory programmed to automatically

diagnose a fault and to display error codes on the LCD.

PRINTER ERROR

Whenever a printer fault condition is detected, the printer stops and
the message "Printer error" scrolls across the upper LCD with the error
code (see below) fixed on the secondary one.

1 = 

Motor locked

= Printer clutch jammed

= Selection lever fault

I

2

C BUS ERROR

In case of an I

2

C bus fatal error due for example to a defective

EEPROM or RTC, the message "Serial bus error" keeps scrolling across
LCD from right to left indefinitely. Meter should be returned for repair
(see warranty section).

FAULT CONDITIONS

Содержание HI 98230

Страница 1: ...Instruction Manual HI 98230 HI 98240 Printing and Logging pH Meters with GLP and Smart Electrodes w w w h a n n a i n s t c o m...

Страница 2: ...meters are also equipped with a stability indicator and backlight feature for comfortable reading even in dark environments The GLP features provide a guarantee of data consistency Measurements can b...

Страница 3: ...the LOBAT indication is displayed to warn the user of this condition However the meter continues to measure correctly even when the low battery indication is displayed When the batteries become too we...

Страница 4: ...lectrode remove the pH electrode from the sample before turning the meter off If the meter is OFF detach the electrode from the meter before immersing the electrode in the storage solution ON OFF SPEC...

Страница 5: ...iption of the specific setup items their valid values and the factory settings default Code Valid values Default 00 Lot data printing scrolling 00 to 16 00 01 Print lots data summary 02 Printer enable...

Страница 6: ...sample is reached press and hold the ALT and PAPER keys until the printer stops During printout the LCD will display the sample number being Code Valid values Default 10 Show pH GLP data 11 Show ORP G...

Страница 7: ...age GLP data printing will scroll across LCD The meter will then print a complete set of the following GLP data a Instrument identification ID code printed at that moment If printout is stopped the LC...

Страница 8: ...ferent samples it is recommended to rinse the electrode thoroughly with deionized water or if not available tap water first and then with some of the next sample to condition the electrode before imme...

Страница 9: ...ORP electrode to the meter and press ON OFF to power on the instrument The meter automatically sets the mV range If the ORP electrode is not connected the LCD will display the no probe message and th...

Страница 10: ...ng will be printed beside the date Note A dashed line indicates that the calibration has never been performed 6 If password is correct or not set the meter will display 7 01pH on the up per LCD and b1...

Страница 11: ...when LCD is displaying mV 3 User is prompted to enter password if it has been set to a value different from 0000 otherwise skip to step 6 4 Enter the password with the arrow keys 5 Press ALT and CFM t...

Страница 12: ...ly enabled in order to return to logging with printing mode LOG ON DEMAND In measuring or Timed log mode press LOG to store the current reading The LCD will display Two different modes to print log da...

Страница 13: ...the meter on again CALIBRATION ALARM TIME OUT The calibration alarm time out is available only for pH calibration It is possible to set through setup code 13 the number of days before the next requir...

Страница 14: ...t is possible to make any elaboration available with software e g graphics statistic analysis HI 92000 offers a variety of features and is provided with an on line help to support the user throughout...

Страница 15: ...lot was the 16th the new logging period restarts from lot 01 overwriting previously logged data When Timed logging memory is full the meter overwrites the oldest lot data progressively reducing the o...

Страница 16: ...I 98240 can also be used with other adapters In this case remember to check the correct polarity of your adapter before connecting it to the meter BATTERY REPLACEMENT TEMPERATURE RESISTANCE CORRELATIO...

Страница 17: ...own could be due to Clogged Dirty Junction refer to the Cleaning Procedure above Loss of shielding due to low electrolyte level in refillable electrodes only refill with fresh HI 7071 for single junct...

Страница 18: ...7 01 buffer solution 500 mL FDA bottle HI 8009L pH 9 18 buffer solution 500 mL FDA bottle HI 7009L pH 9 18 buffer solution 500 mL bottle HI 7010L pH 10 01 buffer solution 500 mL bottle HI 8010L pH 10...

Страница 19: ...10005 Voltage adapter 115 Vac 12 Vdc HI 710006 Voltage adapter 230 Vac 12 Vdc HI 710031 Rugged carrying case HI 710034 Paper roll 10 pcs HI 710035 Ink cartridge HI 76405 Electrode holder HI 8427 pH mV...

Страница 20: ...s in workmanship and materials when used for their intended purpose and maintained according to instructions The electrodes and the probes are warranted for a period of six months This warranty is lim...

Страница 21: ...42 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 0...

Отзывы: