• buffer value, with sign and decimal point and expo-
nent (11 chars).
• calibration time,
yymmddhhmmss
(12 chars).
• electrode condition, with sign (3 chars). The “-01”
code means not calculated.
• Rel mV calibration data (if available), which contains:
• the calibration offset, with sign (7 chars)
• the calibration time,
yymmddhhmmss
(12 chars).
• ISE calibration data (if available), which contains:
• the number of calibrated standards (1 char)
• the ion charge, with sign (2 chars)
• the calibration slope, with sign and decimal point (7 chars)
• the calibration time,
yymmddhhmmss
(12 chars)
• standards information (for each standard)
• type (1 char): 0 - always standard solution.
• status (1 char):N (new) - calibrated in last calibration;
O (old) - from an old calibration.
• warnings during calibration (2 chars): 00 - no warning.
• standard value, with sign and decimal point and
exponent (11 chars).
• calibration time,
yymmddhhmmss
(12 chars).
PAR
Requests the setup parameters setting.
The answer string contains:
• Instrument ID (4 chars)
• Calibration Alarm time out for pH (2 chars)
• Calibration Alarm timeout for ISE (2 chars) - if ISE
available
• SETUP information (2 chars): 8 bit hexadecimal encoding.
• 0x01 - beep ON (else OFF)
• 0x04 - degrees Celsius (else degrees Fahrenheit)
• 0x08 - Offset calibration (else Point calibration)
• Auto-off/Light time (3 chars)
• Auto power off time (3 chars)
• The number of custom buffers (1 char)
• The custom buffer values, with sign and decimal point,
for each defined custom buffer (7 chars)
• The ID of the ISE electrode (2 chars) - if ISE available
• The molar weight of the selected ION, with sign and
decimal point (9 ASCII characters)