17
• Press
CAL
. The instrument will display the measured pH, the LCD first
expected buffer and the temperature reading.
• If necessary, press the
ARROW
keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer,
CFM
functional
key is displayed.
• Press
CFM
to confirm first point.
• The calibrated value and the second expected buffer value is then dis-
played on the LCD.
• After the first calibration point is confirmed, immerse the pH electrode and the
temperature probe approximately 4 cm (1½”) into the second buffer solution
and stir gently. The temperature probe should be close to the pH electrode.
• If necessary, press the
ARROW
keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.