26
• Press
CAL
when the instrument is in
RELATIVE mV
measurement mode.
The relative mV value and the temperature values are displayed.
• Use the
ARROW
keys if you want to change the displayed relative mV value.
• When the reading is stable, in mV range and the Relative mV offset is
inside the offset window (±2000 mV),
CFM
functional key is displayed
.
• Press
CFM
to confirm relative mV calibration. The instrument will return to
measurement mode.
• If the absolute mV reading is out of range or the Relative mV offset is out
of the offset window,
“Wrong relative offset”
message is displayed.
Change the input value or the Relative mV value to complete the calibra-
tion process.