This prompt appears for 1 second each time the instrument is turned
on.

This indicates that the meter is in a ready state and zeroing can be
performed.

Sampling in Progress. This flashing prompt appears each time the
meter is performing a measurement.

This indicates that the meter is in a zeroed state and measurement
can be performed.

Light over range: the cuvet is not inserted correctly and an excess

ambient light is reaching the detector. If the cover is properly installed,
then contact your dealer or the nearest Hanna Customer Service
Center.

The blinking "BAT" indicates that the battery voltage is getting low
and the battery needs to be replaced.

This indicates that the battery is dead and must be replaced. Once
this indication is displayed, the meter will lock up. Change the

battery and restart.

“Configuration”: this indicates that the meter has lost its configuration.
Contact your dealer or the nearest Hanna Customer Service Center.

CALIBRATION MODE  MESSAGES

The date of the last calibration appears on the LCD each time the

calibration mode

 is entered. If calibration is performed for the first

time, “F.CAL” appears.

This indicates that the factory calibration is selected.

GUIDE TO DISPLAY CODES

12

• It is important that the sample does not contain any

debris. This would corrupt the reading.

• Each time the cuvet is used, the cap must be

tightened to the same degree.

• Whenever the cuvet is placed into the measurement

cell, it must be dry outside, and completely free of
fingerprints, oil or dirt. Wipe it thoroughly with

HI 731318

 (tissue for wiping cuvets, see chapter

ACCESSORIES) or a lint-free cloth prior to insertion.

• Shaking the cuvet can generate bubbles in the

sample, causing higher readings. To obtain accurate
measurements, remove such bubbles by swirling or
by gently tapping the vial.

• Do not let the reacted sample stand too long after

reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row, but

it is recommended to take a new zero reading for
each sample and to use the same cuvet for zeroing
and measurement.

• After the reading it is important to discard

immediately the sample, otherwise the glass might

become permanently stained.

• All the reaction times reported in this manual are

referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved
at 30°C (86°F).

• In order to maximize accuracy, prior to a

measurement follow the 

validation procedure

 to

be sure that the instrument is properly calibrated. If

necessary, calibrate the instrument.

Содержание HI 95701C

Страница 1: ...1 Instruction Manual HI 95701C Free Chlorine ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95701R2 05 06 w w w h a n n a i n s t c o m...

Страница 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Страница 3: ...ER ACCESSORIES HI 95701 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass...

Страница 4: ...HI 95701 meter measures the free chlorine Cl2 content in water samples in the 0 00 to 5 00 mg L ppm range The method is an adaption of the Standard Method 4500 Cl G for drinking water The reagent is i...

Страница 5: ...a standard solution of 1 00 mg L chlorine and a representative lot of reagent an operator obtained with a single instrument a standard deviation of 0 02 mg L SPECIFICATIONS PRECISION AND ACCURACY 16 C...

Страница 6: ...narrow band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When...

Страница 7: ...vet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal D...

Страница 8: ...icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCEDURE Turn the meter on by pressi...

Страница 9: ...bration is selected GUIDE TO DISPLAY CODES 12 It is important that the sample does not contain any debris This would corrupt the reading Each time the cuvet is used the cap must be tightened to the sa...

Страница 10: ...ed thesampleandthezerocuvetareinverted Zero a zero reading was not taken Follow the instruction in the measurement procedure for zeroing the meter 11 The instructions listed below should be carefully...

Отзывы: