Range

0.00 to 5.00 mg/L

Resolution

0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L

Precision

±0.02 mg/L @ 1.00 mg/L

Typical EMC Deviation

±0.01 mg/L

Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Light Detector

Silicon Photocell

Method

Adaptation of the Standard Method 4500-Cl G. The reaction between
free chlorine and the DPD reagent causes a pink tint in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in 

measurement mode

;

after 1 hour of non-use in 

calibration mode

.

Dimensions

180 x 83 x 46 mm (7.1 x 3.3 x 1.8")

Weight

290 g (10 oz.).

REQUIRED REAGENTS
Code

Description

Quantity/test

HI 93701-0

Free Chlorine Reagent

1 packet

Precision is how closely repeated
measurements agree with each other.
Precision is usually expressed as standard
deviation (SD).
Accuracy is defined as the nearness of a test

result to the true value.
Although good precision suggests good
accuracy, precise results can be inaccurate.
The figure explains these definitions.
In a laboratory using a standard solution of
1.00 mg/L chlorine and a representative lot
of reagent, an operator obtained with a single
instrument a standard deviation of
0.02 mg/L.

SPECIFICATIONS

PRECISION AND ACCURACY

16

CALIBRATION PROCEDURE

• Turn the meter on by pressing ON/OFF.
• When the LCD displays “701”, it is ready.

• Enter the 

calibration mode

 by holding CAL CHECK

for three seconds.

• The date of the last calibration appears (e.g.:

month “01”, day “08”). “F.CAL” means that the
factory calibration is selected.

Note

: at this point it is possible to reset the

instrument to restore FACTORY CALIBRATION, see
instructions page 18.

• Place the CAL CHECK™ Standard Cuvet 

A

 into

the holder and ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO and “SIP” will blink on the display.

• After a few seconds the display will show “-0.0-”.

The meter is now zeroed and ready for calibration.

• Remove the cuvet.

• Place the CAL CHECK™ Standard HI 95701-11

Cuvet 

B

 into the holder and ensure that the notch

on the cap is positioned securely into the groove.

Note

:

 

it is possible to interrupt calibration procedure at any time by pressing ON/OFF.

Warning

: do not validate or calibrate the instrument with standard solutions other than the Hanna

CAL CHECK™ Standards, otherwise erroneous results will be obtained.

or

CALIBRATION PROCEDURE

Содержание HI 95701C

Страница 1: ...1 Instruction Manual HI 95701C Free Chlorine ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95701R2 05 06 w w w h a n n a i n s t c o m...

Страница 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Страница 3: ...ER ACCESSORIES HI 95701 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass...

Страница 4: ...HI 95701 meter measures the free chlorine Cl2 content in water samples in the 0 00 to 5 00 mg L ppm range The method is an adaption of the Standard Method 4500 Cl G for drinking water The reagent is i...

Страница 5: ...a standard solution of 1 00 mg L chlorine and a representative lot of reagent an operator obtained with a single instrument a standard deviation of 0 02 mg L SPECIFICATIONS PRECISION AND ACCURACY 16 C...

Страница 6: ...narrow band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When...

Страница 7: ...vet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal D...

Страница 8: ...icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCEDURE Turn the meter on by pressi...

Страница 9: ...bration is selected GUIDE TO DISPLAY CODES 12 It is important that the sample does not contain any debris This would corrupt the reading Each time the cuvet is used the cap must be tightened to the sa...

Страница 10: ...ed thesampleandthezerocuvetareinverted Zero a zero reading was not taken Follow the instruction in the measurement procedure for zeroing the meter 11 The instructions listed below should be carefully...

Отзывы: