SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
HI 92701 AND HI 92704
HI 92701 AND HI 92704
HI 92701 AND HI 92704
HI 92701 AND HI 92704
HI 92701 AND HI 92704
HI 92701C
HI 92701F
HI 92704C
HI 92704F
Range
-200.0 to 999.9°C
-300.0 to 999.9°F
1000 to 1370°C
1000 to 2500°F
Resolution
0.1/1°C
0.2/1°F
Accuracy
±0.3% of reading
±0.3% of reading
(@25°C/77°F)
or ±0.5°C
or ±1°F
for one year, excluding probe error
Typical EMC Deviation
±5°C
±10°F
Power Supply
4x1.5V AA size, 500 hours with 60' printing
interval. Socket for 12VDC adapter
Auto Shut-off
After 5 minutes of non-use
Channels
1 for HI 92701; 4 for HI 92704
Printer
Low-power impact belt type, 14 characters per line
using 38 mm plain paper
Printing Intervals
1, 2, 5, 10, 15, 30, 60, 120 and 180 minutes
Environment
0 to 50 °C (32 to 122°F) Max. 95% RH
Dimensions
220 x 82 x 66 mm (8.7 x 3.2 x 2.6")
Shipping
Instrument: 500g (18 oz.)
Weight
Kit: 1.4 kg (3.1 lb.)
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
HI 92801 AND HI 92804
HI 92801 AND HI 92804
HI 92801 AND HI 92804
HI 92801 AND HI 92804
HI 92801 AND HI 92804
HI 92801C
HI 92801F
HI 92804C
HI 92804F
Range
-200.0 to 999.9°C
-300.0 to 999.9°F
1000 to 1370°C
1000 to 2500°F
Resolution
0.1/1°C
0.2/1°F
Accuracy
±0.3% of reading
±0.3% of reading
(@25°C/77°F)
or ±0.5°C
or ±1°F
for one year, excluding probe error
Typical EMC Deviation
±5°C
±10°F
Power Supply
4x1.5V AA size, 500 hours with 60' printing
interval. Socket for 12VDC adapter
Auto Shut-off
After 5 minutes of non-use
Channels
1 for HI 92801; 4 for HI 92804
Printer
Low-power impact belt type, 14 characters per line
using 38 mm plain paper
Printing/Logging Intervals
1, 2, 5, 10, 15, 30, 60, 120 and 180 minutes
Environment
0 to 50 °C (32 to 122°F) Max. 95% RH
Dimensions
220 x 82 x 66 mm (8.7 x 3.2 x 2.6")
Shipping
Instrument: 500g (18 oz.)
Weight
Kit: 1.4 kg (3.1 lb.)
Untitled-27
8
8
ON/OFF
TEMP
INITIAL PREPARATION
Each meter is supplied complete with batteries. Remove the back
cover, unwrap the batteries and install them while paying attention
to the polarity.
To prepare the instrument for use, connect a K-type thermocouple
temperature probe to the connector located on the top of the
instrument.
To switch the
HI92701
and the
HI92704
on,
press the ON/OFF key.
To switch the
HI92801
and the
HI92804
on,
press the TEMP key.
Choose the most appropriate temperature probe(s) for your applica-
tion (see page 30) and connect it (them) to the meter.
HI 92701
and
HI 92801
have a single probe connection.
HI 92704
and
HI 92804
have four probe connectors.
With the meter facing you, channel #1 is the first connector on the
top left hand side.
If the temperature probe is not connected to the instrument, the
meter will display and print "----" to alert the user.
This also indicates the possibility of a broken probe cable. The channel
is displayed with "CH" and the number of channel (four channels
versions only).
Note:
To ensure accurate readings it is recommended that the
temperature probes be connected to consecutive channels be-
ginning with channel #1. For example, if only one probe is to
be used, it must be connected to channel #1. If 2, 3 or 4
probes are to be used, they must be connected to channels 1,
2, 3, and 4 in ascending order.