Instruction Manual

HI9060
HI9061

HI91541

Portable Waterproof

Microprocessor

Thermistor

Thermometers

ins

e

rum

n s

CLEAR

HI

LO

HOLD

HI 9060

microcomputer

thermometer

°C/°F

°C

ON/OFF

-50 .. .. .. .. +150°C

-58 .. .. .. .. +302°F

h t t p : / / w w w . h a n n a c a n . c o m

These Instruments are in

Compliance with the CE Directives

Содержание HI 9061

Страница 1: ...able Waterproof Microprocessor Thermistor Thermometers ins e rum ns CLEAR HI LO HOLD HI 9060 microcomputer thermometer C F C ON OFF 50 150 C 58 302 F h t t p w w w h a n n a c a n c o m These Instrume...

Страница 2: ...cise idea of its versatility in a wide range of applications These instruments are in compliance with directives EN 50081 1 and 50082 1 TABLE OF CONTENTS Preliminary Examination 3 General Description...

Страница 3: ...meters are very easy to use in any field where high humidity levels and dust could effect the performance of a conventional thermometer HI91541 provides accurate and consistent readings not only for l...

Страница 4: ...HI764 calibration test plugs are also available to check the meter s accuracy see page 13 HI9060 can measure in Centigrade and in Fahrenheit Other features of HI9060 include display and clearing of th...

Страница 5: ...ife 4 x 1 5V AA 200 hours Environment 10 to 50 C 14 to 122 F 100 RH Dimensions 196 x 80 x 60 mm 7 7 x 3 1 x 2 4 Weight 425 g 15 oz FUNCTIONAL DESCRIPTION HI 9060 CLEAR HI LO HOLD HI 9060 microcomputer...

Страница 6: ...luding probe error Typical EMC 0 5 C 1 F Deviation with HI 765BL probe Probe HI765 series optional Battery Type Life 4 x 1 5 V AA 200 hours Environment 0 to 50 C 32 to 122 F 100 RH Dimensions 196 x 80...

Страница 7: ...ear excluding probe error Typical EMC 0 5 C 1 F Deviation with HI 764 probe Probe HI764 series optional Battery Type Life 4 x 1 5 V AA 200 hours Environment 0 to 50 C 32 to 122 F 100 RH Dimensions 196...

Страница 8: ...in their place Each time the thermometer is switched on it will carry a self test to ensure that no segment is missing The LCD shows a full set of fig ures during the self test After the test the ther...

Страница 9: ...made by pressing the respective HI or LO keys The appropriate display indicator will appear together with the value retrieved from memory Release the key to resume normal operation By pressing the CLE...

Страница 10: ...inking M on the display indicates the operating mode Press the same key again to resume normal opera tion MEASURING SCALE C F for HI9060 only When the meter is turned on it automatically defaults to t...

Страница 11: ...and HI9061 stop to operate when the low battery signal is detected The user is warned by the code Eb HI91541 displays BAT when the batteries are run down See also the Battery Replacement section on pa...

Страница 12: ...s out of range Eb for HI9060 Battery weak need to be and HI9061 replaced BAT for HI91541 All segments lit Initial input falls out of in for more than 3 strument range For HI 9060 only C The instrument...

Страница 13: ...ers by more than 0 4 C 0 8 F from the Test Plug value the unit is due for recalibration Choose the right Test Plug to suit your needs For HI9060 and HI9061 HI765 18C Calibration key 18 0 C 0 4 C HI765...

Страница 14: ...ust only take place in a non hazardous area using the battery types specified in this instruction manual see page 17 In order to replace rundown batteries simply remove the two screws on the rear cove...

Страница 15: ...switch from one to an other without wasting time and money going through tedious calibration procedures They are available in different colors to avoid cross contamination during testing For HI 9060 a...

Страница 16: ...th 1 m 3 3 cable and white handle HI764A 10 Air probe with 10 m 33 cable and white handle HI764L General purpose liquid probe with 1 m 3 3 cable and white handle HI764L 10 General purpose liquid probe...

Страница 17: ...s Temperature Probes 3 mm 0 12 100 mm 3 9 205 mm 8 1 HI764P HI765P 3 mm 0 12 100 mm 3 9 205 mm 8 1 3 mm 0 12 100 mm 3 9 205 mm 8 1 3mm 0 12 HI764A HI765A HI764L HI765L HI764W HI765W Other Accessories...

Страница 18: ...and the nature of the failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned...

Страница 19: ...ACor60VDC Toavoiddamagesorburns donotperformanymeasurementinmicrowaveovens InparticularcasesHI9060couldswitchfrom Cto Forvice versa Insuchcasespressthe C F keytoresumethepropermeasurementrange Inparti...

Страница 20: ...of applications The reference literature currently covers areas such as Water Treatment Process Swimming Pools Agriculture Food Laboratory Thermometry and many others New reference material is constan...