The
HI 88713
Turbidity bench meter has a built in complete GLP information. The calibration date
and the calibration points are displayed in a comprehensive mode for each range.
To display the GLP information, simply press
CAL
key. A screen with instrument serial number and
with information about the calibration is displayed. For further information, press the “
GLP
”
functional key.
The GLP contains:
•
Instrument serial number
•
The last user calibration date, in
selected format and time in hh.mm format. If no
calibration was performed, the “Not calibrated”
message is displayed and the meter uses only the
factory calibration.
•
Parameter as NTU Ratio Turbidity, NTU
Non Ratio Turbidity, FAU Turbidity or FNU Turbidity.
•
The value of each calibration point (up to
5 points for NTU ratio mode; 4 points for FNU, FAU
and NTU non-ratio modes). If the first calibration
point was skipped, the 0.00 value is displayed.
GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
To restore the factory calibration for the currently
selected range, press
CAL
key when in the main
screen. The first screen for the GLP is displayed.
Press
“Delete”
functional key to initiate the
calibration delete procedure and then press
“CFM”
functional key to delete the user calibration and
restore the Factory calibration.
Note:
Only the user calibration for the current
selected range is deleted.