4
A B B R E V I A T I O N S
A B B R E V I A T I O N S
A B B R E V I A T I O N S
A B B R E V I A T I O N S
A B B R E V I A T I O N S
NTU
Nephelometric Turbidity Units
FAU
Formazin Attenuation Units
FNU
Formazin Nephelometric Units
ISO
International Organization for
Standardization
LCD
Liquid Crystal Display
RTC
Real Time Clock
RH
Relative Humidity
ID
Identification
EBC
European Brewery Comitee
Turbidity is the optical property that causes light to be scattered and absorbed, rather than transmitted.
The scattering of the light as it passes through a liquid is caused by the suspended solids. The higher the
turbidity, the greater the amount of scattered light. The molecules in a very pure fluid scatter light to a
certain degree, no solution will have zero turbidity.
The
ISO 7027 Method
specifies the key parameters for the optical system and the method to
measure the turbidity of water.
The
HI 88713
Turbidity bench meter is designed to meet or exceed the criteria specified by the
ISO
7027, DIN 38404
and
NF EN 27027
.
The light beam that passes through the sample is scattered in all directions. The intensity and pattern
of the scattered light is affected by many variables including the wavelenght of the incident light,
particle size and shape, refractive index and color.
The optical system of the
HI 88713
turbidity bench meter uses an IR LED, a scattered light detector
(90
0
) and a transmitted light detector (180
0
).
PRINCIPLE OF OPERATION
PRINCIPLE OF OPERATION
PRINCIPLE OF OPERATION
PRINCIPLE OF OPERATION
PRINCIPLE OF OPERATION
The
HI 88713
Turbidity bench meter features G.L.P. (Good Laboratory Practice) functions that allow
traceability of the calibration conditions. The last calibration points as well as time and date
The
HI 88713
Turbidity bench meter has a user-friendly interface with an easy to understand,
graphical LCD. Comprehensive contextual help is available at a simple key press. All messages and
help screens are available in several languages. Furthermore, a tutorial mode guides the user step by
step through the analysis process.
The HI 88713
HI 88713
HI 88713
HI 88713
HI 88713 features a logging function that can hold up to 200 measurements in the internal
memory. Data can be viewed directly on the screen or it can be downloaded to a PC for further
analysis using the HI9200 Windows
®
compatible software.