Note
When adjusting values using the
or
keys it is important to allow for sufficient
response time (up to 30 seconds)
The table below lists the values of output codes along with the calibration point values
(which are the analog output minimum and the analog output maximum) as indicated
on the display.
The secondary display indicates the current calibration point value, while primary display
indicates the current calibration type.
OUTPUT
CALIBRATION CALIBRATION CALIBRATION
TYPE
CODE POINT1 POINT2
0-1 mA
0
0 mA
1 mA
0-20 mA
1
0 mA
20 mA
4-20 mA
2
4 mA
20 mA
0-5 Vdc
3
0 Vdc
5 Vdc
1-5 Vdc
4
1 Vdc
5 Vdc
0-10 Vdc
5
0 Vdc
10 Vdc
• Wait for approximately 30 seconds (until the reading of the calibrator is stable).
• Press CFM to confirm. The meter will switch to the second calibration point. Repeat
the above procedure.
• After the desired readings are obtained, press CFM and the meter will skip back to
normal operational mode.