38
Note
If the entered cell constant value is invalid, the “ERROR” indicator blinks on the LCD.
Note
Press SETUP before CFM to exit without changing the cell constant.
CALIBRATION BUFFER DIRECT SELECTION
This feature allows to set a user-defined calibration point, in order to perform calibration
at a point different from the memorized standards.
• Press CAL to enter calibration mode. The LCD will show 0.
•
Press SETUP key.
•
Using
,
and
,
enter the desired buffer value and confirm by pressing CFM.
Note
Press SETUP before CFM to exit without changes.
Note
It is suggested to calibrate the offset before entering the calibration buffer direct selection.
TEMPERATURE CALIBRATION
The controller is factory calibrated for temperature. However, the user may also perform
a one point temperature calibration. This procedure is to calibrate the offset only; the
slope will remain as factory calibrated.
• Prepare a beaker containing a solution at a given temperature inside the range of
the meter.
• Use a Checktemp or a calibrated thermometer with a resolution of 0.1° as a reference
thermometer.
• Immerse the temperature probe in the beaker as
close to the Checktemp as possible.
• Press and hold first CFM and then CAL to enter the
temperature calibration mode.
• Execute the password procedure.
• Select code 1 via the arrow keys for temperature calibration and confirm with CFM.