Code
Valid Values
Default PW
30 Relay 3 high alarm (HA)
0.5 to 99.5% full scale
HA-Hys≥A+Hys,Hys=1.5%f.s.
HA≥S1 or HA≥S2
95%
f.s.
no
31 Relay 3 low alarm (LA)
0.5 to 99.5% full scale
LA+Hys+HA-Hys
Hys=1.5%fs,LA+S1 or LA+S2
5% f.s. no
32 Proportional control
mode period
1 to 30 min
5
no
33 Maximum relay ON time
(after which an alarm
mode is entered)
10 to 9999 min
60
no
34 Alarm mask time
00:00 to 30:00
00:00
no
40 Analog output selection
0: 0-1mA
1: 0-20 mA
2: 4-20 mA
3: 0-5 VDC
4: 1-5 VDC
5: 0-10 VDC
2
no
41 Analog output
lower limit (O_VARMIN)
0 to 100% full scale
(O_VARMIN T O_VARMAX-5% f.s.)
0
no
42 Analog output
upper limit (O_VARMAX)
0 to 100% full scale
(O_VARMIN T O_VARMAX - 5% f.s.)
100%
f.s.
no
60 Current day
01 to 31
from
RTC
no
61 Current month
01 to 12
from
RTC
no
62 Current year
1998 to 9999
from
RTC
no
63 Current time
00:00 to 23:59
from
RTC
no
71 Baud rate (RS485)
1200, 2400, 4800, 9600
9600
no
72 Cleaning timer
0 to 19999 days
0
no
73 Initial cleaning day
01 to 31
01
no
74 Initial cleaning month
01 to 12
01
no