89
• If an offset calibration has been
made, the instrument will turn to
“slope” calibration (as indicated
by the “SLO” message on the
secondary display . The slope value
is shown on the primary LCD and
the first digit is blinking to permit
modifications) .
• Press the
or
key to modify the value or
key to
move to the next digit .
• Once selected the desired value, press CFM to confirm .
• After confirmation the instrument will turn back to normal
mode .
• If the slope is invalid, the ”WRONG” indicator will blink
on the LCD .
Note
Press LCD or CAL DATA key to exit calibration . The slope
will be set to the default value (57 .5 mV/pH) .
After direct selection of offset and slope, calibration data
can be viewed by pressing the CAL DATA key while in normal
mode, but no pH-buffer values will be displayed during the
data scrolling on the LCD .