background image

Conmet

®

1

 (HI 98305) 

• 

Conmet

®

2

 

 (HI 98306) 

• 

Conmet

®

3

 (HI 98307)

Conductivity Meters with Automatic Temperature Compensation

ISTCONMET   05/05

Specifications

Conmet

®

1

Conmet

®

2

Conmet

®

3

Range

0 to 1999 µS/cm

0.00 to 19.99 mS/cm

0.00 to 19.99 ppt (g/L)

Resolution

1  µS/cm

0.01 mS/cm

0.01 ppt (g/L)

Accuracy (@20°C/68°F)

±2% F.S.

Calibration

Manual, 1 point, through slope trimmer

Temp. Compensation

Automatic, 0 to 50°C, with ß=1.9% ±0.15% /°C, and 25°C (77°F) reference temperature

Probe (included)

HI 3291

HI 3292

HI 3292

Battery Type / Life

4 x 1.5V alkaline / approx. 150 hours of continuous use

Environment

0 to 50°C (32 to 122°F); 95% RH max.

Dimensions / Weight

265 x 29 x 15 mm (10.4 x 1.1 x 0.6’’) / 400 g (0.88 lb.) complete with the carrying case

Each meter is supplied in a rugged plastic carrying case with 4-ring conductivity probe, calibration solution bottle (120 mL), screwdriver and instructions.

Operations: • Connect the conductivity probe to the meter (Fig. 1) and turn the unit on by sliding the ON/OFF switch on the top of the instrument.

• Dip the probe into the solution by ensuring that all 4 rings of the probe are immersed and without exceeding the maximum level (Fig. 2).
• Stir gently to remove any air bubbles trapped inside the probe and wait for the reading to stabilize.
• To minimize errors, it is recommended to rinse the probe with some of the sample to be tested before taking measurement.
• After use, switch the unit off, rinse the probe in alcohol for a few minutes.
• If the display fades, replace the batteries. Switch the unit off, pull out the battery compartment and replace the batteries (Fig. 3).
  Battery replacement must only take place in a non hazardous area using the battery type specified in this instruction manual.

Calibration: • Dip the probe into the proper (supplied) calibration solution and wait for the reading to stabilize.

• If necessary, correct the reading to the solution value by slowly turning the calibration trimmer with the supplied screwdriver.

These meters are in
compliance with the

CE directives