4. Operation
If the device lid cannot be opened, you can activte the emergency open manually.
4.12 Emergency Lid Open
1. Find the emergency lid-lock release hole in the front body of the device and take
out the white rubber closure
2. Insert the emergency lid open tool into the hole and revolve it counter clockwise until
the lid is released.
· Wait for the rotor to stop completely before activating the emergency open.
· After opening the lid manually, it is recommended to wait until normal
electricity comes back.