6.4.3 If the bowl cap sealing gasket is damaged
·
During operation, a leak occurs in the bowl cap due to rotational force, resulting in leakage of the
injected sample. Please replace it immediately.
·
If a large scratch or separation phenomenon is found on the contact surface, please replace it
immediately.
[Caution] When replacing, always check whether the bowl cap is positioned correctly.
6.4.4 Rubber coupling is damaged
①
After checking that the rubber coupling is in close contact with the inside of the pulley cap, replace it
immediately if it is cracked or damaged.
: Remove and replace cracked or damaged rubber coupling inside the pulley cap.
rubber coupling
pulley cap
rubber coupling
pulley cap
②
Visually check the female & male clutch and replace it immediately if damage is confirmed.
6.4.5 If the drag bushing is damaged
①
When checking with the naked eye, if the inner diameter is large, please replace it.
②
If the degree of wear is biased to one side, please check the device balance before replacement.
HANIL Scientific Inc.
│
ihanil.com
│
Service: +82-2-3452-8966 / [email protected]
56