USER MANUAL
│
J1250
-
RPM/RCF
The rotation speed is expressed in RPM/RCF, and the maximum RPM is 15,000 and the
maximum RCF is 15,801 xg. Also, the RPM/RCF value of the bowl is automatically
calculated and displayed as a set value (set speed).
-
TIME
The time can be set in hours, minutes and seconds, and can be set up to a maximum of
99 hours 59 minutes 59 seconds.
-
ACC/DEC
The steps for setting the acceleration/deceleration time of the rotor can be set by dividing it
into 3 acceleration steps from 1 to 3 and 4 deceleration steps from 0 to 3.
The higher the number, the faster the step.
-
Start / Stop
It is used to start and stop motion.
-
Pump
The pump operation status is displayed, and the icon is activated during operation.
The basic operation is that when the set RPM is reached, the pump automatically turns on
and runs, and stops when it stops. In case of manual control during operation, ON/OFF
control can be performed by touching the Pump Icon
(Optional) When adding a level sensor
- When Level Sensor 1 is detected, a warning message is displayed, and when Level 2
Sensor is detected, the pump is automatically turned off.
-
Door
Displays Door Open/Closed status. In the Door Open state, even if Start is executed, it
does not operate, and when Door Open is detected during operation, it is automatically
stopped with an error.
-
SET
You can check the setting value.
-
Program
You can set program setting, loading, and deletion. Up to
99 programs can be saved.
HANIL Scientific Inc.
│
ihanil.com
│
Service: +82-2-3452-8966 / [email protected]
25
Содержание J1250
Страница 1: ...User Manual J1250 Continuous Tubular Centrifuge...
Страница 3: ...USER MANUAL J1250 HANIL Scientific Inc ihanil com Service 82 2 3452 8966 techsupport ihanil com 3...