Mount the Drag body assembly using Urethane hammer (refer to p48)
·
Carefully insert the bowl into the drag bushing hole in the frame. (see p48)
·
Install the Supernatant Outlet and Supernatant Outlet lid. (see p49)
·
Close the main body door. After applying maximum tension to the door lock handle device by
hand, tap it 2-3 times with a Urethane hammer to finish.
·
Lower the spindle (refer to page 50)
·
After confirming that the lower part of the spindle and the upper groove of the bowl are aligned,
fasten the coupling nut and bowl using a coupling wrench and urethane hammer. (see p50)
·
Pull down the spindle protective sleeve and tighten it as much as possible. (see p51)
·
Connect the Sample-in tube to the sample inlet and connect it to the sample bottle through the
pump head of the sample feeding pump.
·
Connect the supplied outlet hose to the supernatant outlet on the top of the device to collect the
centrifugation supernatant. (See p13 & p53)
·
Turn the grease cup halfway and put grease in to lubricate the bowl rotation.
- It is recommended to rotate the grease cover 1/4 turn every 60 minutes during operation.
Grease
Grease cup: Lubricating grease injection device when bowl rotates
Grease
⑪
Check No. 1 to No. 8 again and balancing with reference to p21~p22 once again to check
the state of the device and, if it is correctly joined, operate the motor.
HANIL Scientific Inc.
│
ihanil.com
│
Service: +82-2-3452-8966 / [email protected]