6. Troubleshooting
6.2 Error messages
Error
Display
Cause
Solution
No display
None
- No power connection
- Blackout
- Check the power connection
- Check the main fuse in
laboratory and device
No running
Unknow
Rotor ID!
- No rotor
- Errors in drive or rotor
recognition function
- Load a rotor
- Restrat the device
-Check the rotor sensor for
foreign substance.
Hi
Temperature
Error!
-The current temperature
is higher than the set
temperature
- Set the temperature higher
Low
Temperature
Error!
-The current temperature
is lower than the set
temperature.
- Set the temperature lower
The lid is not
open
None
-Blackout
- Emergency lid open
The lid is not
closed
To close the
door
- No working of lid pin
- No power connection
- Check the power connection
- Close the lid
On
acceleration,
the device is
vibrating
** Imbalance
Error! **
-Rotor is loaded
asymmetrically
- Not tighten the rotor
with shaft
- Load the rotor symmetrically
- Tighten the rotor to the
motor shaft.
Failure of
power supply
** Power
Failure! **
- Failure of power supply
- check the outlet
Etc
** Motor
Overheat! **
- Motor overheat
- Stop the centrifugation
- Let the device cool down
** Over
Speed! **
-The current speed is
higher than the set speed
- Stop the centrifugation
- Repeat the run