background image

8

Version 2.1 

 

 

 

 

 

 

 

 

 

 

        Display 

        S21 Mainswitch 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S7 Footrestswitch   

 

 

 

S8 Footrestswitch 

S9 Footrestswitch 

 

 

 

 

 

 

 

 

Powered swivel 

Motor 

S25 Seat swivel switch  

Board P3 

S34 Emergency stop switch 

 

 

 

 

 

S7/8 

S7/8 

S9 

S25 

S34 

Display
S21 Mainswitch

S7 Footrestswitch
S8 Footrestswitch
S9 Footrestswitch

Powered swivel
Motor
S25 Seat swivel switch 
Board P3
S34 Emergency stop switch

Version 2.1 

 

 

 

 

 

 

 

 

 

 

        Display 

        S21 Mainswitch 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S7 Footrestswitch   

 

 

 

S8 Footrestswitch 

S9 Footrestswitch 

 

 

 

 

 

 

 

 

Powered swivel 

Motor 

S25 Seat swivel switch  

Board P3 

S34 Emergency stop switch 

 

 

 

 

 

S7/8 

S7/8 

S9 

S25 

S34 

Version 2.1 

 

 

 

 

 

 

 

 

 

 

        Display 

        S21 Mainswitch 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S7 Footrestswitch   

 

 

 

S8 Footrestswitch 

S9 Footrestswitch 

 

 

 

 

 

 

 

 

Powered swivel 

Motor 

S25 Seat swivel switch  

Board P3 

S34 Emergency stop switch 

 

 

 

 

 

S7/8 

S7/8 

S9 

S25 

S34 

Содержание 2000

Страница 1: ...Handicare 2000 QUICK REFERENCE GUIDE VERSION 2 1...

Страница 2: ...2...

Страница 3: ...Display codes 3 3 References 5 4 Glossary 6 5 Testing 10 6 Display 11 7 Toggle 11 8 The remotes 11 9 Safety edges 12 10 The batteries 13 11 Charging power supply 14 12 Boards 15 13 Options 16 14 Rescu...

Страница 4: ......

Страница 5: ...tested where the components are located and how they react to certain situations The subsequent chapters deal with repairing faults Flow charts are used in some cases where step by step choices have t...

Страница 6: ...Low battery voltage 31 8 UP travel direction 21 9 DOWN travel direction 21 A Hinge open 35 b Toggle switch active at power up 21 C IR address fail 32 d Relay not open 20 E Relay not closed 20 F Brake...

Страница 7: ...voltage 31 8 UP travel direction 21 9 DOWN travel direction 21 A Hinge open 35 b Toggle switch active at power up 21 C IR address fail 32 d Relay not open 20 E Relay not closed 20 F Brake semi conduct...

Страница 8: ...limit switch 22 27 28 35 The batteries 24 31 34 The charger 31 35 The tractionmotor 21 The brake 21 The display 24 The on off switch 24 The toggle 23 26 The armrestswitches 23 26 The hinge 33 34 35 T...

Страница 9: ...7 18 Final limit switch S10 Speed switch lower trunnion S22 Hinge enable switch top trunnion not visible S13 S14 S17 S18 S15 S16 S10 S13 14 15 16 Limit switch S 17 18 Final limit switch S10 Speed swit...

Страница 10: ...runion Safety brake S19 S20 Overspeed governor switch Infrared receiver 5 Amp fuse 30 Amp fuse Main board P1 Motor Brake S11 S12 S35 S11 12 Front chassis edge switch S35 Handwinderswitch Contact charg...

Страница 11: ...itch Powered swivel Motor S25 Seat swivel switch Board P3 S34 Emergency stop switch Version 2 1 9 Display S21 Mainswitch S7 Footrestswitch S8 Footrestswitch S9 Footrestswitch Powered swivel Motor S25...

Страница 12: ...0 Armrestswitches S31 Foldable hinge contact S32 Foldable hinge contact S33 Foldable hinge contact Motor Board P4 S29 30 S23 S24 S33 S31 S32 S24 Toggle S23 Keyswitch S29 S30 Armrestswitches S31 Foldab...

Страница 13: ...rrying out continuity testing Set the multimeter to the beep or resistance or voltage test mode Check that you are testing between the correct connector pins numbers No beep or resistance means that t...

Страница 14: ...a special button for parking the lift On the backside is a red orange or green cover Green means the remote functions on isplay ft is a display which shows a diagnostic code to help you to locate the...

Страница 15: ...osite direction to remove the block The diagnostic code shows 4 or 6 The following switches have been included in the direction sensitive safety edge Footrest switches Detection switches for the power...

Страница 16: ...eries is at least 5V less than that of the other it can be assumed the battery is faulty If the voltage between the batteries varies there is something wrong with them Guarantee No guarantee is provid...

Страница 17: ...ift is powered off then back on but as soon as lift docks on contacts process will repeat again This fault code is linked to fault code 7 but the code 7 is not usually seen as code 1 takes priority Co...

Страница 18: ...nboard P1 126 Powered swivel P3 Powered footrest raiser P2 126 Foldable hinge P4 12 Boards 125 Mainboard P1 126 Powered swivel P3 Powered footrest raiser P2 126 Foldable hinge P4 125 Mainboard P1 126...

Страница 19: ...oking for different parameters to determine when and where power should be applied to the attached motor When the parameters are met the MS125 pcb passes the battery voltage through the communication...

Страница 20: ...the down direction After a short pause the hinge will lower On the rail at the top and just above the foldable hinge are 2 contacts The contact on the upper tube is for the communication and the cont...

Страница 21: ...h the user on the lift when a safety measure is switched off If it is not possible to remove the fault without danger to the user first release the user In such case ensure that your are positioned ab...

Страница 22: ...charge See page 31 Remote con trol problems See page 32 Powered swi vel problems See page 36 Hinge operated but lift does not move Code J See page 34 The hinge does not operate See page 33 Powered foo...

Страница 23: ...r current Code L Current limit exceed Problem with brake Problem with display Board P1 is faulty Main board isfaulty Check motor and track for obstructions There is a break in the current between M13...

Страница 24: ...s switch is switched on Code Flat line See page 23 Code 0 See page 22 Main board is faulty Main board is faulty Main board is faulty Code 8 or 9 flashing There is a break in the current between M1 and...

Страница 25: ...tions don t function either There is a break in the current between M8 7 U8 1 and U18 1 Short circuit between safety edge and frame There is a break in the current between M8 9 U8 2 U10 2 U11 2 U11 1...

Страница 26: ...and ground is 4 5 V The voltage between M7 1 or M7 3 and ground is not 4 5 V when calling the lift The main board P1 is faulty The lift is on the end position go to Lift does not move in 1 direction...

Страница 27: ...none The lift has run for 30 seconds and the thermal fuse on the main board is hot Main switch S21 switched off Faulty brake The main switch S21 is faulty Problem with infrared receiver Broken fuse 2...

Страница 28: ...Quick Reference Guide Handicare 2000 25 The lift does not move in one direction Code 5 See page 28 Code 4 See page 29 Code 3 See page 27 Code 6 See page 30 Code flat line See page 26...

Страница 29: ...3 and ground is 4 5 V When moving the toggle The main board P1 is faulty There is break in the current between M7 1 U1 3 U2 3 U6 3 U12 3 U13 3 There is break in the current between M7 3 U1 1 U2 1 U6...

Страница 30: ...oltage between M8 5 and ground is 3 5 V The voltage between M8 6 and ground is not 3 5 V Main board P1 is faulty Main board P1 is faulty Connect the molex testing link The voltage between M8 5 and gro...

Страница 31: ...and ground is 3 5 V The voltage between M8 4 and ground is not 3 5 V Main board P1 is faulty Main board P1 is faulty Connect the molex testing link The voltage between M8 3 and ground is not 3 5 V Th...

Страница 32: ...tside of the stairs The lift will not move up The voltage between M7 5 and ground is not 3 5 V There is an obstacle blocking the upperside of the pack or the upperside of the footrest There is a break...

Страница 33: ...The lift will not move up The voltage between M7 8 and ground is not 3 5V There is an obstacle blocking the upperside of the pack or the upperside of the footrest There is a break in the current in th...

Страница 34: ...cted 33V PSU faulty Faulty contact strip in the unit The charging power supply is not con nected to the mains Break in the current between the contact strip and M10 Break in the current from the charg...

Страница 35: ...Problems with the infrared receiver The thermal fuse on the board is hot The infrared receiver is faulty There is a break in the current between the infrared receiver and M5 1 2 3 or M5 1 2 3 Code C T...

Страница 36: ...33v There is a break in the current between M3 and M12 The 5 Amp fuse is faulty There is a break in the current between the contact strip and M12 There is a bad contact between the contact in the pack...

Страница 37: ...but not on the top stop The lift thinks he is at the hinge point Adjust the top stop Check the hinge activation bracket and the roller switch S22 on the lift There is a stopping difference between th...

Страница 38: ...n M23 1 and M23 2 or M23 3 and M23 4 The top endswitch of the folding hinge S33 is faulty If the top limit switch is defective after a short delay the lift would be allowed to drive UP but the hinge w...

Страница 39: ...3 is wrong The key switch S23 is faulty There is a break in the ground connection M18 The power has not been turned off on after changing the settings 10 sec delay The main board P1 is faulty Board P3...

Страница 40: ...switch S21 is faulty The keyswitch S23 is faulty There is a break in the current between U15 or U17 and U7 Broken fuse 25 A There is a break in the current between M8 9 U8 2 U10 2 U11 2 U11 1 U10 1 an...

Страница 41: ...runs on half speed Code n The batteries are low There is a problem with the magnets on the rail The reedcontact S10 is faulty There is a break in the current between M20 1 and M20 4 The main board P1...

Страница 42: ...Quick Reference Guide Handicare 2000 39...

Страница 43: ......

Страница 44: ......

Отзывы:

Похожие инструкции для 2000