I m p o r t a n t   h i n t s          

class 2. The instrument may be used in any position, however, 

sufficient ventilation must be assured as convection cooling is 

used. For continuous operation prefer a horizontal or slightly 

upward position using the feet.

STOP

TiPP

  Do not cover either the holes of the case nor the 

cooling fins. 

Specifications with tolerances are valid after a 30 minute warm-

up period and at 23 degrees C. Specifications without tolerances 

are typical values of an average instrument.

Warranty and Repair

HAmeG instruments are subjected to a rigorous quality control. 

Prior to shipment each instrument will be burnt in for 10 hours. 

Intermittent operation will produce nearly all early failures. 

After burn in, a final functional and quality test is performed 

to check all operating modes and fulfilment of specifications. 

The latter is performed with test equipment traceable to na-

tional measurement standards.

Statutory warranty regulations apply in the country where the 

HAmeG product was purchased. In case of complaints please 

contact the dealer who supplied your HAmeG product. 

Maintenance

The instrument does not require any maintenance. Dirt may 

be removed by a soft moist cloth, if necessary adding a mild 

detergent. (Water and 1 %.) Grease may be removed with ben-

zine (petrol ether). Displays and windows may only be cleaned 

with a moist cloth.

 

Do not use alcohol, solvents or paste. Under no cir-

cumstances any fluid should be allowed to get into 

the instrument. If other cleaning fluids are used 

damage to the lacquered or plastic surfaces is pos-

sible. 

Mains voltage

A main voltage of 115V and 230V can be chosen. Please check 

whether the mains voltage used corresponds with the volta-

ge indicated by the mains voltage selector on the rear panel. 

If not, the voltage has to be changed. In this case the line fuse 

has to be changed, too.

 

Please note:

 

After changing the main voltage, the line fuse has 

to be changed. Otherwise the instrument may be 

destroChanging the line fuse

Changing the line fuse

The fuses are accessible from the outside and contained in 

the  line  voltage  connector  housing.  Before  changing  a  fuse 

disconnect the instrument from the line, the line cord must 

be removed. Check fuse holder and line cord for any damages. 

Use a screw driver to loosen the fuse holder screw counter-

clockwise while pressing the top of the fuse holder down. The 

top holding the fuse will then come off. exchange the defective 

fuse against a correct new one.

It is forbidden to repair defective fuses or to bridge them by any 

means. Any damage caused this way will void the warranty.

          I m p o r t a n t   h i n t s

Types of fuses: 

   

 

Size 5 x 20 mm; 250V~,

IeC 60127-2/5

eN 60127-2/5

Line voltage 

 

Correct fuse type

230 V          

 

2 x 1.6 A slow blow

115 V           

 

2 x 3.2 A slow blow

STOP

TiPP

Содержание HM8143

Страница 1: ...A r b i t r a r y P o w e r S u p p l y H M 8 1 4 3 Manual 99 Washington Street Melrose MA 02176 Fax 781 665 0780 TestEquipmentDepot com...

Страница 2: ...uence of such signals is unavoidable This will not cause damage or put the instrument out of operation Small deviations of the measuring value reading exceeding the instruments speci cations may resul...

Страница 3: ...ies 23 Linear power supplies 23 Switched mode power supplies SMPS 23 Parallel and series operation 23 Current limit 24 Electronic fuse 24 Introduction to the operation 24 Operation modes 24 Constant v...

Страница 4: ...ion 10 mV 1 mA Arbitrary waveform power supply 1024 points 12 bit Tracking mode for 30 V outputs External modulation of output voltages Electronic load up to 60 W per channel max 2 A SENSE lines Multi...

Страница 5: ...ry Function Channel I only Number of points 1024 Resolution 12 Bit Parameters of points Dwell time and Voltage Dwell time 10 s 60 s Repetition rate 1 255 and continuous Inputs Modulation input BNC soc...

Страница 6: ...nd guarantee operation without any danger to the operator According to safety class 1 requirements all parts of the housing and the chassis are con nected to the safety ground terminal of the power co...

Страница 7: ...e petrol ether Displays and windows may only be cleaned with a moist cloth Do not use alcohol solvents or paste Under no cir cumstances any fluid should be allowed to get into the instrument If other...

Страница 8: ...it CURRENT pushbutton and LED Setting of current limit via frontpanel By pushing the but ton the setting function is active and the appertaining LED is lit Rotary knob Parameter setting of voltage and...

Страница 9: ...ut voltage is recti ed smoothed and then chopped The capacitance values needed here for ltering the 100 resp 120 Hz ripple are higher due to the lower voltage All SMPS feature a very much higher ef ci...

Страница 10: ...STOP STOP Introduction to the operation First time operation Please observe especially the following notes The line voltage indicated on the rear panel corresponds to the available line voltage also t...

Страница 11: ...ermit easy triggering of an oscilloscope connected to the output of the HM8143 especially in arbitrary mode the in strument is equipped with a BNC socket TRIGGER IN OUT on its rear panel This is con g...

Страница 12: ...instrument by short cir cuits or overheating Current limit If one of the output voltages is short circuited the current li miter automatically keeps the current from rising beyond the programmed maxim...

Страница 13: ...I2 0 123 A RU1 RU2 Format RU1 or RU2 Reply U1 12 34V or U2 12 34V Function The voltage values sent back by the HM8143 are the programmed voltage values Use the MUx com mands to query the actual values...

Страница 14: ...nal proceeds internal This also has the effect of switching off the outputs Pressing this button again switches the outputs of the power supply on The waveform generation starts either after the recei...

Страница 15: ...epeat this sequence 10 times The required data table is as follows ABT A10 00_B30 00_A30 00_725 67_002 00_002 00_N10 or ABT A10 00_B30 00_A30 00_725 67_002 00_002 00_N10 2 x 100 s 2 00V 100 ms 25 67 V...

Отзывы: