Compensation of the voltage drop across the cables

Power source

Sense error

amplifier

 source
variable

+ Source

cable

+ Sense Pick-Up

– Source

cable

– Sense Pick-Up

load

Connecting the load

The load has to be connected to the middle safety terminals  , 

 or 

. For the connection please use 4 mm banana plugs.

STOP

  Please note the polarity of the load terminals: the 

red terminal is the positive, the black terminal is 
the negative connector.

The transparent terminals 

/

 are the SENSE inputs. With 

these SENSE terminals the voltage loss across the cables 
can be compensated. The HM8143 balances this voltage loss 
automatically and the load will see the voltage set. Connect 
two separate measurement cables in parallel to the connec-
ting cables of the load.

          O p e r a t i o n  o f  t h e  H M 8 1 4 3

Operation of the HM8143

Setting output voltages and the current limits

The changeable parameters (output voltages and current li-
mit) are set using the rotary knob

. To change values, fi rst 

select the appropriate parameter with the VOLTAGE

 

/

 and 

CURRENT

 

 

 buttons. Then use the rotary knob

 to set 

the desired value.

If the outputs are on (OUTPUT LED 

 is on) the HM8143 dis-

plays will show the actual values, that means the power supply 
will show the measured values of voltage and current (Vout and 
Iout). Operating the VOLTAGE 

/

 or the CURRENT button 

/

 will switch the HM8143 to setting mode, which is being indi-

cated by glowing of  one of the LEDs above the buttons VOLTAGE 

/

 or CURRENT 

/

. The corresponding display will show 

the nominal value of the output voltage or current limit. Now 
the desired value of the output voltage or current limit can be 
adjusted with the rotary knob

. This mode will be left after 

about 2 seconds after the last operation of the rotary knob. The 
HM8143 will then display the measured values of the output 
voltage and current again.

Trigger Input + Trigger Output (Start/Stop)

In order to permit easy triggering of an oscilloscope connected 
to the output of the HM8143, especially in arbitrary mode, the in-
strument is equipped with a BNC socket TRIGGER IN/OUT 

 on 

its rear panel. This is confi gured as a tristate output and permits 
a trigger signal to be taken after each signal period in arbitrary 
mode, or the arbitrary function to be activated by an external 
trigger signal (TTL level).

Modulation inputs

By virtue of the modulation inputs  MODULATION R/L 

 on 

the rear panel of HM8143, it  can be also be used as a modula-
tion power amplifi er. The input voltage is amplifi ed with factor 
3. The frequency range (-3 dB) goes from DC to 50 kHz. The 
allowable external voltage ranges from 0 V to 10 V.

The output voltage of HM8143 will be the sum of:
 
V

out

  = (V

modin

 x 3) + V

set

!

  Please note that the sum V

out

 = (V

modin

 x 3) + V

set

 

must not exceed the value of 30 V, as then the pro-
per functionality of the current regulation  is not 
ensured and the connected load can be destroyed.

Electronic load

The  HM8143  also  offers  a  mode  in  which  it  functions  as  an 
electronic load (current sink). The instrument goes into this 
mode automatically, and it can be recognized by a negative sign 
(–) in front of a displayed current value. The same limit values 
apply to voltage and current as in normal operating mode. In 
this operation mode the output voltage measured is normally 
greater than the nominal value (V

actual

 

>

 V

set

)

Serial or parallel operation

To increase the output voltages and currents, the two chan-
nels of the power supply can be connected either in series or 
in parallel.

 

It is important to keep in mind that when the two 
output circuits are connected in series a greater 
voltage than that ordinarily permitted for safety 
reasons can develop. The HM8143 may therefore 
be used only by personnel who are familiar with 
the associated risks.

Arbitrary waveform mode

By interface the HM8143 can also be made to generate freely 
programmable waveforms within the limit values set (arbitrary 
mode). See chapter Arbitrary.

Example:   Modulationsource:   V

mod

   = 2,0 V

ss

 

 

f

mod

    = 50 Hz

 channel 

I  V

set

  

1

0 V

 channel 

II  V

set

    = 10 V 

0 V

1 V

2 V

modulation signal

Содержание HM8143

Страница 1: ...A r b i t r a r y P o w e r S u p p l y H M 8 1 4 3 Manual 99 Washington Street Melrose MA 02176 Fax 781 665 0780 TestEquipmentDepot com...

Страница 2: ...uence of such signals is unavoidable This will not cause damage or put the instrument out of operation Small deviations of the measuring value reading exceeding the instruments speci cations may resul...

Страница 3: ...ies 23 Linear power supplies 23 Switched mode power supplies SMPS 23 Parallel and series operation 23 Current limit 24 Electronic fuse 24 Introduction to the operation 24 Operation modes 24 Constant v...

Страница 4: ...ion 10 mV 1 mA Arbitrary waveform power supply 1024 points 12 bit Tracking mode for 30 V outputs External modulation of output voltages Electronic load up to 60 W per channel max 2 A SENSE lines Multi...

Страница 5: ...ry Function Channel I only Number of points 1024 Resolution 12 Bit Parameters of points Dwell time and Voltage Dwell time 10 s 60 s Repetition rate 1 255 and continuous Inputs Modulation input BNC soc...

Страница 6: ...nd guarantee operation without any danger to the operator According to safety class 1 requirements all parts of the housing and the chassis are con nected to the safety ground terminal of the power co...

Страница 7: ...e petrol ether Displays and windows may only be cleaned with a moist cloth Do not use alcohol solvents or paste Under no cir cumstances any fluid should be allowed to get into the instrument If other...

Страница 8: ...it CURRENT pushbutton and LED Setting of current limit via frontpanel By pushing the but ton the setting function is active and the appertaining LED is lit Rotary knob Parameter setting of voltage and...

Страница 9: ...ut voltage is recti ed smoothed and then chopped The capacitance values needed here for ltering the 100 resp 120 Hz ripple are higher due to the lower voltage All SMPS feature a very much higher ef ci...

Страница 10: ...STOP STOP Introduction to the operation First time operation Please observe especially the following notes The line voltage indicated on the rear panel corresponds to the available line voltage also t...

Страница 11: ...ermit easy triggering of an oscilloscope connected to the output of the HM8143 especially in arbitrary mode the in strument is equipped with a BNC socket TRIGGER IN OUT on its rear panel This is con g...

Страница 12: ...instrument by short cir cuits or overheating Current limit If one of the output voltages is short circuited the current li miter automatically keeps the current from rising beyond the programmed maxim...

Страница 13: ...I2 0 123 A RU1 RU2 Format RU1 or RU2 Reply U1 12 34V or U2 12 34V Function The voltage values sent back by the HM8143 are the programmed voltage values Use the MUx com mands to query the actual values...

Страница 14: ...nal proceeds internal This also has the effect of switching off the outputs Pressing this button again switches the outputs of the power supply on The waveform generation starts either after the recei...

Страница 15: ...epeat this sequence 10 times The required data table is as follows ABT A10 00_B30 00_A30 00_725 67_002 00_002 00_N10 or ABT A10 00_B30 00_A30 00_725 67_002 00_002 00_N10 2 x 100 s 2 00V 100 ms 25 67 V...

Отзывы: