29
Subject to change without notice
R e m o t e c o n t r o l
6
h
=
50 ms
7
h
=
100
ms
8
h
=
200 ms
9
h
=
500 ms
A
h
= 1
s
B
h
= 2
s
C
h
= 5
s
D
h
=
10
s
E
h
=
20 s
F
h
=
50
s
Example: It is wished to program the following waveform:
1
s
10.00
V
3
s
30.00
V
100 ms
25.67 V
200 us
2.00 V
It is also wished to repeat this sequence 10 times.
The required data table is as follows:
ABT:A10.00_B30.00_A30.00_725.67_002.00 _002.00_N10 or
ABT A10.00_B30.00_A30.00_725.67_002.00_002.00_N10
2 x 100 μs 2.00V
100 ms 25.67 V
2 s+1 s 30.00 V
1 s 10.00 V
RUN/STP
Format: RUN
Function: Starts waveform generation in ARB mode
Format: STP
Function : Interrupts the arbitrary function while running.