Line Impedance Stabilization Network  HM6050-2

– Maximum continuous current: 16A
– Phase indication via LED on front panel
– BNC test signal output, L1 or N
– Artificial  Hand
– Protective earth simulation circuit
– Remote control (EIA-232 interface)
– Transient  Limiter

The HM6050-2 Line Impedance Sta-

bilization Network (LISN) meets standards
VDE 0876 and CISPR Publ. 16. It contains air
core inductance coils and features an
Artificial Hand and a PE simulating network
which can be bridged.

Conducted emissions on AC power lines,

which are typically generated by electrical
equipment, can be verified with the help of a
LISN together with a Spectrum Analyzer/EMC
Receiver. The DUT (device under test) must
be connected directly to the LISN. Inside the
LISN the power lines are terminated with a
well-defined impedance network, against
each other and against ground.

The HM6050-2 Line Impedance Sta-

bilization Network (LISN) in principle is a
filter network. Through a low pass filter the

DUT is connected to the AC power lines.
The LISN also presents a well-defined
impedance to the signal. For measurements
with a Spectrum Analyzer/EMC Receiver
the EMC signal is available after having
passed a high pass filter.

Two identical networks provide the asym-

metric noise emission signals of the DUT’s
power lines L1 and N. The user can choose
between the signals, the selected one will
be available at the HM6050-2’s test signal
outlet. The stabilization network (simulation
for the AC power lines) is arranged in form
of a "V".

When working with a Spectrum Analyzer/

EMC Receiver it is highly recommended to
enable the built-in Transient Limiter of the
HM6050-2.

Hameg HM6050-2D Shown

Содержание HM6050-2

Страница 1: ...d a n c e S t a b i l i z a t i o n N e t w o r k H M 6 0 5 0 2 Manual 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com Hameg H...

Страница 2: ...50 2 20 Specifications 21 Operating Manual 22 Introduction 22 Used Symbols 22 Operation according to rules 22 Safety 22 Replacement of fuses 23 Operating conditions 24 Warranty and repair 24 Operating...

Страница 3: ...uction in the manual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters If an interface has several connectors only one connector must have a connection to a...

Страница 4: ...WG EMC Directive 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amend e par 91 263 EWG 92 31 CEE Directriz EMC 89 336 CEE enmendada por 91 263 CE 92 91 CEE Niederspannungsrichtlin...

Страница 5: ...tly to the LISN Inside the LISN the power lines are terminated with a well defined impedance network against each other and against ground The HM6050 2 Line Impedance Sta bilization Network LISN in pr...

Страница 6: ...output with PE Test signal output 50 BNC jack Test Receiver Rear panel Power line cable 3 x 1 5 mm2 male with PE Ground Alu Block with tread 4 mm 40 x 20 x 20 mm RS 232 Interface 9 pin Sub D female Tr...

Страница 7: ...ns It conforms to the regulations as defined in CISPR Publ 16 or EN55011 Safety This instrument has been designed and tested in accordance with VDE 0411 Part 1 Safety requirements for electrical equip...

Страница 8: ...M6050 2 contains a fuse Ex factory the HM6050 2 is set to 230V 10 50 60Hz If the instrument shall operate at a mains voltage of 115 V 10 the wire bridges have to be changed by soldering Insert or repl...

Страница 9: ...ly failures are detected After burn in a thorough test of all functions and of quality is run all specifications and operating modes are checked In case of reclamations during the two years warranty p...

Страница 10: ...nd connection according to Class 1 2 Ground jack This jack is used for testing DUTs which don t have a protective ground line in their AC power cable but do have a separate ground connection The DUT s...

Страница 11: ...tached Test Receiver s or Spectrum Analyzer s input circuitry from high transient voltages After pushing button 10 the transient limiter circuit will be bypassed A blinking red LED 12 indicates this o...

Страница 12: ...onnected between the ground block and the chamber s shielding material VDE 0877 Part1 16 RS 232 interface Safety hints Attention All interface lines are galvanic coupled to the LISN As mentioned in pa...

Страница 13: ...upper case letters Command Function Indication R Remote control enabled RM LED lit O Local control RM LED dark P Protect ground line sim circuit on PE LED lit p Protect ground line sim circuit off byp...

Страница 14: ...hich may be touched by the user lie on a dangerous high voltage level EMC measurement procedures for LISN HM6050 2 Measurements have to be performed according to the VDE regulations VDE 0877 Part 1 Af...

Отзывы: