background image

Subject to change without notice

29

according to IEC364-4-41 (CENELEC HD 384.4.41 and VDE 0100 Part
410). Due to the circuitry of the HM6050-2 a maximum leakage current
of 800 mA can occur.

It is not possible to operate the HM6050-2 at a mains outlet with
a leakage current circuit breaker.

Attention! Danger of death!
The LISN has to be grounded according to the regulations (Cenelec
HD384/DIN VDE0100) separately; otherwise conductive parts of the
housing, which may be touched by the user, lie on a dangerous
high voltage level.

EMC measurement procedures for LISN HM6050-2

Measurements have to be performed according to the VDE regulations
VDE 0877 Part 1.

After the test setup is completed according to the regulations one can
begin with the measurement procedures.

Follow these procedure steps:

1. Set HM6050-2 to operation (power on)

(Transient limiter in active status (LED off))

2. Set DUT (Equipment under test) to operation
3. Set Spectrum Analyzer/Test Receiver to operation
4. Connect HM6050-2 to the input jack of the Spectrum Analyzer/Test

Receiver via a coaxial HF cable.

Visit us at www.TestEquipmentDepot.com

Содержание HM6050-2

Страница 1: ...d a n c e S t a b i l i z a t i o n N e t w o r k H M 6 0 5 0 2 Manual 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com Hameg H...

Страница 2: ...50 2 20 Specifications 21 Operating Manual 22 Introduction 22 Used Symbols 22 Operation according to rules 22 Safety 22 Replacement of fuses 23 Operating conditions 24 Warranty and repair 24 Operating...

Страница 3: ...uction in the manual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters If an interface has several connectors only one connector must have a connection to a...

Страница 4: ...WG EMC Directive 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amend e par 91 263 EWG 92 31 CEE Directriz EMC 89 336 CEE enmendada por 91 263 CE 92 91 CEE Niederspannungsrichtlin...

Страница 5: ...tly to the LISN Inside the LISN the power lines are terminated with a well defined impedance network against each other and against ground The HM6050 2 Line Impedance Sta bilization Network LISN in pr...

Страница 6: ...output with PE Test signal output 50 BNC jack Test Receiver Rear panel Power line cable 3 x 1 5 mm2 male with PE Ground Alu Block with tread 4 mm 40 x 20 x 20 mm RS 232 Interface 9 pin Sub D female Tr...

Страница 7: ...ns It conforms to the regulations as defined in CISPR Publ 16 or EN55011 Safety This instrument has been designed and tested in accordance with VDE 0411 Part 1 Safety requirements for electrical equip...

Страница 8: ...M6050 2 contains a fuse Ex factory the HM6050 2 is set to 230V 10 50 60Hz If the instrument shall operate at a mains voltage of 115 V 10 the wire bridges have to be changed by soldering Insert or repl...

Страница 9: ...ly failures are detected After burn in a thorough test of all functions and of quality is run all specifications and operating modes are checked In case of reclamations during the two years warranty p...

Страница 10: ...nd connection according to Class 1 2 Ground jack This jack is used for testing DUTs which don t have a protective ground line in their AC power cable but do have a separate ground connection The DUT s...

Страница 11: ...tached Test Receiver s or Spectrum Analyzer s input circuitry from high transient voltages After pushing button 10 the transient limiter circuit will be bypassed A blinking red LED 12 indicates this o...

Страница 12: ...onnected between the ground block and the chamber s shielding material VDE 0877 Part1 16 RS 232 interface Safety hints Attention All interface lines are galvanic coupled to the LISN As mentioned in pa...

Страница 13: ...upper case letters Command Function Indication R Remote control enabled RM LED lit O Local control RM LED dark P Protect ground line sim circuit on PE LED lit p Protect ground line sim circuit off byp...

Страница 14: ...hich may be touched by the user lie on a dangerous high voltage level EMC measurement procedures for LISN HM6050 2 Measurements have to be performed according to the VDE regulations VDE 0877 Part 1 Af...

Отзывы: