Rear panel

(14) AC power cable

The power cable is used to attach the HM6050-2 to the mains outlet.
(

See: “Setting into operation”).

(15) Ground block

The ground block is made of aluminum and fastened to the backside
of the housing by two screws (below the mains cable outlet). Terminal
“PE” may not be used as reference ground because it is connected to
the PE line of the mains cable via a filter. If tests are performed outside
an EMC chamber only a short ground cable may be used. Inside a
chamber the ground cable has to be connected between the ground
block and the chamber’s shielding material (VDE 0877, Part1).

(16) RS-232 interface

Safety hints

Attention!
All interface lines are galvanic coupled to the LISN.

As mentioned in paragraph “Safety” the LISN has to be operated on a
three line mains outlet (with protective ground line). An interface cable
between the RS-232 interface of the LISN and the COM port of a PC
causes a galvanic connection between the two devices. To exclude
negative effects on other safety provisions the PC’s power cable must
be connected to the same mains outlet as the LISN.

Disregarding these safety hints (also see “Safety”) will result in
the loss of HAMEG’s warranty in case of damages. HAMEG is
not liable for any injuries to personnel or any equipment.

Содержание HM6050-2

Страница 1: ...d a n c e S t a b i l i z a t i o n N e t w o r k H M 6 0 5 0 2 Manual 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com Hameg H...

Страница 2: ...50 2 20 Specifications 21 Operating Manual 22 Introduction 22 Used Symbols 22 Operation according to rules 22 Safety 22 Replacement of fuses 23 Operating conditions 24 Warranty and repair 24 Operating...

Страница 3: ...uction in the manual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters If an interface has several connectors only one connector must have a connection to a...

Страница 4: ...WG EMC Directive 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amend e par 91 263 EWG 92 31 CEE Directriz EMC 89 336 CEE enmendada por 91 263 CE 92 91 CEE Niederspannungsrichtlin...

Страница 5: ...tly to the LISN Inside the LISN the power lines are terminated with a well defined impedance network against each other and against ground The HM6050 2 Line Impedance Sta bilization Network LISN in pr...

Страница 6: ...output with PE Test signal output 50 BNC jack Test Receiver Rear panel Power line cable 3 x 1 5 mm2 male with PE Ground Alu Block with tread 4 mm 40 x 20 x 20 mm RS 232 Interface 9 pin Sub D female Tr...

Страница 7: ...ns It conforms to the regulations as defined in CISPR Publ 16 or EN55011 Safety This instrument has been designed and tested in accordance with VDE 0411 Part 1 Safety requirements for electrical equip...

Страница 8: ...M6050 2 contains a fuse Ex factory the HM6050 2 is set to 230V 10 50 60Hz If the instrument shall operate at a mains voltage of 115 V 10 the wire bridges have to be changed by soldering Insert or repl...

Страница 9: ...ly failures are detected After burn in a thorough test of all functions and of quality is run all specifications and operating modes are checked In case of reclamations during the two years warranty p...

Страница 10: ...nd connection according to Class 1 2 Ground jack This jack is used for testing DUTs which don t have a protective ground line in their AC power cable but do have a separate ground connection The DUT s...

Страница 11: ...tached Test Receiver s or Spectrum Analyzer s input circuitry from high transient voltages After pushing button 10 the transient limiter circuit will be bypassed A blinking red LED 12 indicates this o...

Страница 12: ...onnected between the ground block and the chamber s shielding material VDE 0877 Part1 16 RS 232 interface Safety hints Attention All interface lines are galvanic coupled to the LISN As mentioned in pa...

Страница 13: ...upper case letters Command Function Indication R Remote control enabled RM LED lit O Local control RM LED dark P Protect ground line sim circuit on PE LED lit p Protect ground line sim circuit off byp...

Страница 14: ...hich may be touched by the user lie on a dangerous high voltage level EMC measurement procedures for LISN HM6050 2 Measurements have to be performed according to the VDE regulations VDE 0877 Part 1 Af...

Отзывы: