Accessories supplied: Line Cord, Operators Manual, 2 Probes 1:1/10:1

The requirements of modern service and industrial applications can

easily be met with 

HAMEG’s HM304

 

microprocessor

 controlled

oscilloscope. The new 

"AUTO SET"

 capability allows for automatic setup,

where the user need only adjust the focus and intensity controls. All other
settings instantaneously and automatically configure in response to the
input signals when "AUTO SET" is active. The instrument’s intelligence
modifies only those parameters that are necessary for proper signal
presentation. This mode typically generates a display with a period of three
signal repetitions, that has an amplitude of approximately six divisions for
single channel and half of that for dual channel operation. Manual control is
also possible for all settings. A 

PC’s serial port

 can control remote

operation via the 

RS-232

 serial interface. 

LED’s

 indicate measurement

range settings and functions.

A memory base that can store six different customer-defined control

scenarios facilitates repeat measurement tasks. User settings can be
saved and recalled without restriction.

The two vertical (Y) channel amplifier’s 

35MHz

 bandwidth display

signals over 

100MHz

. High frequency trigger circuitry permits synchronization

to signals (having 0.5 division amplitude) beyond 100MHz. The high
resolution timebase allows signal expansion in the 'delay' and 'automatic
trigger after delay' modes by a factor of up to 1,000 times. The HM304 is
equipped with a built-in 

Component Tester

, dual frequency (1 kHz / 1 MHz)

calibration signals and mu-metal CRT shielding. A switching mode power
supply saves both energy and weight with the complete unit weighing only
5.5 kg. (12.1 pounds).

This unit offers attributes, only too easy to get used to, that are beyond

compare to other instruments in this price range. Expert users will rapidly
identify that the HM304’s deluxe suite of features shall be the standard for
future high performance oscilloscopes.

Specifications

Vertical Deflection

Operating modes:

 Channel I or II separate,

both Channels (alternated or chopped),
(Chopper frequency approx. 0.5MHz).

Sum 

or

 difference

 with Ch. I and Ch. II

(both channels invertable).

XY-Mode:

 via channel I and channel II

Frequency range:

 2xDC to 35MHz (-3dB)

Risetime: <10ns. Overshoot max. 1%.

Deflection coefficients:

 14 calibrated steps

from 

1mV/div.

 to 

20V/div.

 (1-2-5 sequence)

with variable 2.5:1 up to 

50V/div.

Accuracy in calibrated position:

1mV/div. to 2mV/div.:  ±5%(0 to 10MHz  (-3dB))
5mV/div. to 20V/div.: ±3%

Input impedance:

 1M

 II 20pF.

Input coupling: DC-AC-GD (ground).
Input voltage: max. 400V (DC + peak AC).

Triggering

Automatic: (peak to peak) 

<20Hz-100MHz

 (

 0.5div.)

Normal with level control: 

DC-100MHz

  (

0.5div.)

Slope:

 positive or negative,

ALT. Triggering;

 LED indicator for trigger action

Sources:

 Channel I or II, CH. I alternating CH II,

line and external

Coupling: 

AC

 (10Hz to 100MHz), 

DC

 (0 to 100MHz),

HF

 (1.5kHz to100MHz), 

LF

 (0 to 1.5kHz)

Active TV-Sync-Separator

 (pos. and neg.)

External:

 

0.3V

pp

 from 30 Hz to 30MHz

2nd triggering (Del. Trig.):

 normal with level

control DC to 100MHz

Horizontal Deflection

Time coefficients:

 22 calibrated steps

from 0.5s/div. to 50ns/div. in 1-2-5 sequence
Accuracy in calibrated position: ±3%.
variable 2.5:1 up to 1.25s/div.,
with X-Mag. x10: 10ns/div. ±5%

Holdoff time:

 variable to approx. 10:1

Delay:

 50ms - 100ns, variable 2.5:1 up to 125ms

Bandwidth X-amplifier:

 0-3MHz (-3dB).

Input X-Amplifier via Channel II,
(sensitivity see Channel II specification)
X-Y phase shift: <3° below 120kHz.

Operation / Control

Auto Set 

(automatic parameter selection)

Manual 

(Front Panel switches)

Memory

 for 6 user-defined parameter settings

Remote control 

with built-in RS-232 interface

Component Tester

Test voltage:

 approx. 8.5V

rms

 (open circuit).

Test current: approx. 7mA

rms

 (shorted).

Test frequency:

 approx. 50Hz

Test connection: 2 banana jacks 4mm´Ø
One test lead is grounded (Safety Earth)

General Information

CRT:

 D14-364GY/123 or ER151-GH/-,rectangular

screen (8x10cm) internal graticule

Acceleration voltage:

 approx 2000V

Trace rotation:

 adjustable on front panel

Calibrator:

 square-wave generator (t

r

 <4ns

)

1kHz/1MHz;

 

Output: 0.2V ±1% and 2V

Line voltage:

 100-240V AC ±10%, 50/60Hz

Power consumption:

 approx. 36 Watt at 50Hz.

Min./Max. ambient temperature: -10°C...+40°C

Protective system:

 Safety class I (IEC1010-1)

Weight: approx. 5.6kg (12.4lbs), color: techno-brown
Cabinet: 

W 285, H 125, D 380 mm (11.1x4.9x14.8 inches)

Lockable tilt handle

Subject to change without notice.

10/95

OSCILLOSCOPES

35MHz Analog-Oscilloscope HM 304

with Auto-Set, Save and  Recall (6 Setup Memories)
Remote control via built-in  RS-232 Interface

Vertical: 2 Channels, 1 mV/div to 20 V/div., Component Tester
Time Base: 0.5 s/div.  to 10 ns/div.; Trigger-after-delay; AlternateTrigger
Triggering: DC to 100 MHz; Automatic Peak to Peak; TV-Sync-Separator

Содержание HM304

Страница 1: ...tructions General Information 5 Symbols 5 Use of tilt handle 5 Safety 5 Operating conditions 5 EMC 6 Warranty 6 Maintenance 6 Protective Switch Off 6 Power supply 6 Type of signal voltage 7 Amplitude...

Страница 2: ...uction in the manual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection...

Страница 3: ...nzt durch 93 68 EWG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized...

Страница 4: ...35MHz 3dB Risetime 10ns Overshoot max 1 Deflection coefficients 14 calibrated steps from1mV div to20V div 1 2 5 sequence with variable 2 5 1 up to 50V div Accuracy in calibrated position 1mV div to 2m...

Страница 5: ...three conductor power cord with protective earthing conductor and a plug with earthing contact The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact T...

Страница 6: ...aintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largely certain that all signals are displayed with the accuracy on...

Страница 7: ...ms Veff have 2 83 times the potential difference in Vpp The relationship between the different voltage magnitudes can be seen from the following figure Voltage values of a sine curve Vrms effective va...

Страница 8: ...e base setting indicated by one of the TIME DIV LED s one or several signal periods or only a part of a period can be displayed The time coefficients are stated in s div when the red sec LED and the 0...

Страница 9: ...hes next to the edge are also not taken into account With very severe transient distortions the rise and fall time measurement has little meaning For amplifiers with approximately constant group delay...

Страница 10: ...oscilloscope by depressing the red POWER pushbutton The instrument will revert to its last used operating mode Except in the case of COMP TESTER mode where a trace appears on the screen if the INTENS...

Страница 11: ...should be adjusted The location of the low frequency compensation trimmer can be found in the probe information sheet Adjust the trimmer with the insulated screw driver provided until the tops of the...

Страница 12: ...channell I to channel II and vice versa after each sweep period In DUAL mode the internal trigger source can be switched over from channel I to channel II and vice versa if the TRIG pushbutton is dep...

Страница 13: ...compared with C short circuited then the test voltage leads the reference voltage and vice versa This applies only in the region up to 90 phase shift Therefore C should be sufficiently large and produ...

Страница 14: ...t when external trigger is used the trigger threshold can be stated as vertical display height in div through which the time base generator starts the display is stable and the trigger LED located in...

Страница 15: ...gnals than the DC coupling because the white noise in the trigger voltage is strongly suppressed So jitter or double triggering of complex signals is avoidable or at least reduced in particular with v...

Страница 16: ...er voltage may have a completely different form from the test signal voltage Triggering is even possible in certain limits with whole number multiples or fractions of the test frequency but only with...

Страница 17: ...following explanation assumes that the trace starts on the left vertical graticule line Photo 1 composite video signal MODE undelayed TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Depre...

Страница 18: ...ted in Dual mode under conditions where DUAL chopped mode is active this display mode is not switched off when time coefficients are being reduced 0 2ms div to 0 05 s div for signal expansion in DEL a...

Страница 19: ...also be made to integrated circuits All these components can be tested in and out of circuit The test principle is fascinatingly simple A built in generator delivers a sine voltage which is applied a...

Страница 20: ...ore testing of transistor amplification is not possible but testing of a single junction is easily and quickly possible Since the test voltage applied is only very low all sections of most semiconduct...

Страница 21: ...ould then be connected to the insulated COMP TESTER socket avoiding hum distortion of the test pattern Another way is a test pattern comparison to an identical circuit which is known to be operational...

Страница 22: ...Subject to change without notice 22...

Страница 23: ...of astigmatism A certain loss of marginal sharpness of the CRT is unavoidable this is due to the manufacturing process of the CRT Symmetry and Drift of the Vertical Amplifier Both of these characteri...

Страница 24: ...eck of the mono channel display is unnecessary it is contained indirectly in the tests above stated Triggering Checks The internal trigger threshold is important as it determines the display height fr...

Страница 25: ...and also the influence of the earths magnetic field which is dependent on the instruments North orientation are corrected by means of the TR potentiometer In general the trace rotation range is asymme...

Страница 26: ...hanged Generally max halving or doubling of this resistance value should be sufficient A too small trigger threshold cause double triggering or premature trigger action due to interference pulses or r...

Страница 27: ...R ckgabe Beschreibung PC Scope Scope PC ID ID Daten CR LF data consits of instrument type manufacturer CR R CR LF remote status and baud rate acceptance TRSTA TRSTA b CR LF query for trigger status d...

Страница 28: ...nter 0 13 CH2 GND AC INV2 ON VALUE Counter 0 13 mv DIV 0000 20V DIV 1101 MODE CT XY A TR CHOP ADD 0 TR SOURCE 00 Y1 01 Y2 1x EXT TB1 x10 0 0 TB A TIME Counter 1 26 50ns DIV 00 bis 0 5s DIV 15hex TB2 0...

Страница 29: ...cy ranges AC 10Hz 100MHz DC 0Hz 100MHz HF 1 5kHz 100MHz LF 0Hz 1 5 kHz TV L to trigger on line sync pulses TV F to trigger on separated frame sync pulses Select SLOPE for the leading slope Sync pulse...

Страница 30: ...control for trace sharpness mechanical knob TR Trace rotation mechanical To align trace with horzontal field potentiometer graticule line Compensates adjustment with influence of Earth s magnetic scr...

Страница 31: ...triggering Not available in combination with ext triggering XY or COMP TESTER modes INPUT CH II Channel II signal input BNC connector Input impedance 1M II 20pF AC DC Selects input coupling of CH II...

Отзывы: