Hameg HM304 Скачать руководство пользователя страница 11

11

Subject to change without notice

of the oscilloscope on the place of work. A centred trace
may not align exactly with the horizontal center line of the
graticule. A few degrees of misalignment can be corrected
by a potentiometer accessible through an opening on the
front panel marked

 TR

.

Probe compensation and use

To display an undistorted waveform on an oscilloscope, the
probe must be matched to the individual input impedance of
the vertical amplifier.

For this purpose a square wave signal with a very fast rise
time and minimum overshoot should be used, as the
sinusoidal contents cover a wide frequency range. The
frequency accuracy and the pulse duty factor are not of such
importance.

The built-in calibration generator provides a square wave signal
with a very fast risetime (<4ns), and switch-selectable
frequencies of approx. 1kHz and 1MHz from two output
sockets below the CRT screen.

As the squarewave signals are used for probe compensation
adjustments, neither the frequency accuracy nor the pulse
duty factor are of importance and therefore not specified.
One output provides 0.2Vpp ±1% (tr <4ns) for 10:1 probes,
and the other 2Vpp for 100:1 probes. When the Y deflection
coefficients are set to 5mV/div., these calibration voltages
correspond to a screen amplitude of 4div.

The output sockets have an internal diameter of 4.9mm to
accommodate the internationally accepted shielding tube
diameter of modern Probes and F-series slimline probes. Only
this type of construction ensures the extremly short ground
connections which are essential for an undistorted waveform
reproduction of non-sinusoidal high frequency signals.

Adjustment at 1kHz

The C-trimmer adjustment (low frequency) compensates the
capacitive loading on the oscilloscope input (approx. 20pF
for the HM304). By this adjustment, the capacitive division
assumes the same ratio as the ohmic voltage divider to ensure
the same division ratio for high and low frequencies, as for
DC. (For 1:1 probes or switchable probes set to 1:1, this
adjustment is neither required nor possible). A baseline par-
allel to the horizontal graticule lines is essential for accurate
probe adjustments. (See also 

„Trace rotation TR“

).

Connect the probes (Types HZ51, 52, 53, 54, or HZ36) to the
CH.I input. One deflection coefficient in the 

VOLTS/DIV

section of channel I must lit. If this is not the case depress
the 

CHI

 pushbutton once and switch off channel II by

depressing the 

CHII

 pushbutton once. Set input coupling 

CH

I

 to DC (LED illuminates) and check that 

GD

 is switched off.

The 

CHI

 deflection coefficient must be 5mV/div., and 

TIME/

DIV.

 should be set to 0.2ms/div., and all variable controls to

CAL.

 position. Plug the the probe tip into the appropriate

calibrator output socket, i.e. 10:1 probes into the 0.2V socket,
100:1 probes into the 2V socket.

incorrect

correct

incorrect

Approximately 2 complete waveform periods are displayed
on the CRT screen. The compensation trimmer should be
adjusted. The location of the low frequency compensation

trimmer can be found in the probe information sheet. Adjust
the trimmer with the insulated screw driver provided, until
the tops of the square wave signal are exactly parallel to the
horizontal graticule lines (see 1kHz diagram). The signal height
should then be 4div. ± 0.16div. (= 4 % (oscilloscope 3% and
probe 1%). During this adjustment, the signal edges will
remain invisible.

Adjustment at 1MHz

Probes HZ51, 52 and 54 can also be HF-compensated. They
incorporate resonance de-emphasing networks (R-trimmer
in conjunction with inductances and capacitors) which per-
mit probe compensation in the range of the upper frequency
limit of the vertical oscilloscope amplifier. Only this
compensative adjustment ensures optimum utilisation of the
full bandwidth, together with constant group delay at the
high frequency end, thereby reducing characteristic transient
distortion near the leading edge (e.g. overshoot, rounding,
ringing, holes or bumps) to an absolute minimum.

Using the probes HZ51, 52 and 54, the full bandwidth of the
HM304 can be utilized without risk of unwanted waveform
distortion.

Prerequisite for this HF compensation is a square wave
generator with fast risetime (typically 4ns), and low output
impedance (approx. 50

), providing 0.2V and 2V at a

frequency of approx. 1MHz. The calibrator output of the
HM304 meets these requirements when the 

CAL.

pushbutton is depressed.

Connect the probe to 

CH.I

 input. Depress the 

CAL.

pushbutton for 1MHz. Operate the oscilloscope as described
under 1kHz but select for 0.2µs/div 

TIME/DIV.

 setting.

Insert the probe tip into the output socket marked 0.2V. A
waveform will be displayed on the CRT screen, with leading
and trailing edges clearly visible. For the HF-adjustment now
to be performed, it will be necessary to observe the rising
edge as well as the upper left corner of the pulse top. The
location of the high frequency compensation trimmer(s) can
also be found in the probe information sheet. These R-
trimmer(s) have to be adjusted such that the beginning of
the pulse is as straight as possible. Overshoot or excessive
rounding are unacceptable. The adjustment is relatively easy
if only one adjusting point is present. In case of several
adjusting points the adjustment is slightly more difficult, but
causes a better result. The rising edge should be as steep as
possible, with a pulse top remaining as straight and horizon-
tal as possible.

incorrect

correct

incorrect

After completion of the HF-adjustment, the signal amplitude
displayed on the CRT screen should have the same value as
during the 1kHz adjustment.
Probes other than those mentioned above, normally have a
larger tip diameter and may not fit into the calibrator outputs.
Whilst it is not difficult for an experienced operator to build a
suitable adapter, it should be pointed out that most of these
probes have a slower risetime with the effect that the total
bandwidth of scope together with probe may fall far below
that of the HM304. Furthermore, the HF-adjustment feature
is nearly always missing so that waveform distortion can not
be entirely excluded.

Содержание HM304

Страница 1: ...tructions General Information 5 Symbols 5 Use of tilt handle 5 Safety 5 Operating conditions 5 EMC 6 Warranty 6 Maintenance 6 Protective Switch Off 6 Power supply 6 Type of signal voltage 7 Amplitude...

Страница 2: ...uction in the manual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection...

Страница 3: ...nzt durch 93 68 EWG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized...

Страница 4: ...35MHz 3dB Risetime 10ns Overshoot max 1 Deflection coefficients 14 calibrated steps from1mV div to20V div 1 2 5 sequence with variable 2 5 1 up to 50V div Accuracy in calibrated position 1mV div to 2m...

Страница 5: ...three conductor power cord with protective earthing conductor and a plug with earthing contact The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact T...

Страница 6: ...aintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largely certain that all signals are displayed with the accuracy on...

Страница 7: ...ms Veff have 2 83 times the potential difference in Vpp The relationship between the different voltage magnitudes can be seen from the following figure Voltage values of a sine curve Vrms effective va...

Страница 8: ...e base setting indicated by one of the TIME DIV LED s one or several signal periods or only a part of a period can be displayed The time coefficients are stated in s div when the red sec LED and the 0...

Страница 9: ...hes next to the edge are also not taken into account With very severe transient distortions the rise and fall time measurement has little meaning For amplifiers with approximately constant group delay...

Страница 10: ...oscilloscope by depressing the red POWER pushbutton The instrument will revert to its last used operating mode Except in the case of COMP TESTER mode where a trace appears on the screen if the INTENS...

Страница 11: ...should be adjusted The location of the low frequency compensation trimmer can be found in the probe information sheet Adjust the trimmer with the insulated screw driver provided until the tops of the...

Страница 12: ...channell I to channel II and vice versa after each sweep period In DUAL mode the internal trigger source can be switched over from channel I to channel II and vice versa if the TRIG pushbutton is dep...

Страница 13: ...compared with C short circuited then the test voltage leads the reference voltage and vice versa This applies only in the region up to 90 phase shift Therefore C should be sufficiently large and produ...

Страница 14: ...t when external trigger is used the trigger threshold can be stated as vertical display height in div through which the time base generator starts the display is stable and the trigger LED located in...

Страница 15: ...gnals than the DC coupling because the white noise in the trigger voltage is strongly suppressed So jitter or double triggering of complex signals is avoidable or at least reduced in particular with v...

Страница 16: ...er voltage may have a completely different form from the test signal voltage Triggering is even possible in certain limits with whole number multiples or fractions of the test frequency but only with...

Страница 17: ...following explanation assumes that the trace starts on the left vertical graticule line Photo 1 composite video signal MODE undelayed TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Depre...

Страница 18: ...ted in Dual mode under conditions where DUAL chopped mode is active this display mode is not switched off when time coefficients are being reduced 0 2ms div to 0 05 s div for signal expansion in DEL a...

Страница 19: ...also be made to integrated circuits All these components can be tested in and out of circuit The test principle is fascinatingly simple A built in generator delivers a sine voltage which is applied a...

Страница 20: ...ore testing of transistor amplification is not possible but testing of a single junction is easily and quickly possible Since the test voltage applied is only very low all sections of most semiconduct...

Страница 21: ...ould then be connected to the insulated COMP TESTER socket avoiding hum distortion of the test pattern Another way is a test pattern comparison to an identical circuit which is known to be operational...

Страница 22: ...Subject to change without notice 22...

Страница 23: ...of astigmatism A certain loss of marginal sharpness of the CRT is unavoidable this is due to the manufacturing process of the CRT Symmetry and Drift of the Vertical Amplifier Both of these characteri...

Страница 24: ...eck of the mono channel display is unnecessary it is contained indirectly in the tests above stated Triggering Checks The internal trigger threshold is important as it determines the display height fr...

Страница 25: ...and also the influence of the earths magnetic field which is dependent on the instruments North orientation are corrected by means of the TR potentiometer In general the trace rotation range is asymme...

Страница 26: ...hanged Generally max halving or doubling of this resistance value should be sufficient A too small trigger threshold cause double triggering or premature trigger action due to interference pulses or r...

Страница 27: ...R ckgabe Beschreibung PC Scope Scope PC ID ID Daten CR LF data consits of instrument type manufacturer CR R CR LF remote status and baud rate acceptance TRSTA TRSTA b CR LF query for trigger status d...

Страница 28: ...nter 0 13 CH2 GND AC INV2 ON VALUE Counter 0 13 mv DIV 0000 20V DIV 1101 MODE CT XY A TR CHOP ADD 0 TR SOURCE 00 Y1 01 Y2 1x EXT TB1 x10 0 0 TB A TIME Counter 1 26 50ns DIV 00 bis 0 5s DIV 15hex TB2 0...

Страница 29: ...cy ranges AC 10Hz 100MHz DC 0Hz 100MHz HF 1 5kHz 100MHz LF 0Hz 1 5 kHz TV L to trigger on line sync pulses TV F to trigger on separated frame sync pulses Select SLOPE for the leading slope Sync pulse...

Страница 30: ...control for trace sharpness mechanical knob TR Trace rotation mechanical To align trace with horzontal field potentiometer graticule line Compensates adjustment with influence of Earth s magnetic scr...

Страница 31: ...triggering Not available in combination with ext triggering XY or COMP TESTER modes INPUT CH II Channel II signal input BNC connector Input impedance 1M II 20pF AC DC Selects input coupling of CH II...

Отзывы: