Hameg HM304 Скачать руководство пользователя страница 19

19

Subject to change without notice

SAVE/RECALL

The instrument contains a non volatile memory in which the
actual instruments settings are stored when the instrument is
switched off. After switching on the oscilloscope and a short
internal check routine the last settings become active again.

The memory mentioned before can also be used by the
operator to save 6 different instrument settings and to recall
them. This relates to all settings with the exception off

INTENS, FOCUS, TR

 (trace rotation), 

DEL. POS./HOLD OFF

and the calibrator frequency pushbutton.

To 

SAVE

 

a particular front panel set up:

Press 

SAVE/RECALL

 pushbutton briefly. Now the trigger

coupling scale has the function of a memory location
indicator. The memory location number is indicated on the
left side of the scale below the S/R marking. After the

SAVE/RECALL

 pushbutton was briefly depressed for the

first time the S/R 1 (DC) LED will blink. The memory
location can now be chosen, selecting it by depressing
one of the NORM (up / down) pushbuttons; this causes
no change in the trigger coupling setting. SAVE the front
panel setting by depressing the 

SAVE/RECALL

 pushbutton

until LED stops blinking. The trigger coupling active before
starting the 

SAVE/RECALL

 procedure (and was  saved)

then becomes active again.

To 

RECALL

 

a front panel set up:

Press 

SAVE/RECALL

 pushbutton briefly. Now the trigger

coupling scale has the function of a memory location
indicator. The memory location number is indicated on the
left side of the scale below the S/R marking. After the

SAVE/RECALL

 pushbutton was briefly depressed for the

first time the S/R 1 

(DC)

 LED will blink. The memory

location can now be chosen, selecting it by depressing
one of the 

NORM

 (up / down) pushbuttons; this causes

no change in the trigger coupling setting. 

RECALL

 the front

panel setting by depressing the 

SAVE/RECALL

 pushbutton

briefly. The LED will stop blinking and the selected set up
is enabled. The LED display will now show  the recalled
trigger coupling mode.

The recalled operating conditions are taken over by the
instrument regardless of whether the mechanical knob
settings coincide or not. If a knob is not in its calibrated
detent (CAL), the LED stops blinking. Turning the knob
then reverts to the actual mechanical setting.

To interrupt a 

SAVE

 

or 

RECALL

 

procedure:

If the 

SAVE/RECALL

 pushbutton was depressed

inadvertently, the procedure can be interrupted by
depressing any electronic selection pushbutton with the
exception of the 

SAVE/RECALL

 and 

NORM

 pushbuttons.

Component Tester

The HM304 has a built-in electronic Component Tester
(COMP. TESTER), which is used for instant display of a test
pattern to indicate whether or not components are faulty.
The 

COMP. TESTER

 can be used for quick checks of

semiconductors (e.g. diodes and transistors), resistors,
capacitors, and inductors. Certain tests can also be made to
integrated circuits. All these components can be tested in
and out of circuit.

The test principle is fascinatingly simple. A built-in generator
delivers a sine voltage, which is applied across the component
under test and a built-in fixed resistor. The sine voltage across
the test object is used for the horizontal deflection, and the
voltage drop across the resistor (i.e. current through test
object) is used for vertical deflection of the oscilloscope. The
test pattern shows a current-voltage characteristic of the test
object.

Since this circuit operates with a frequency of 50Hz (±10%)
and a voltage of approx. 7V max. (open circuit), the indicating
range of the component tester is limited. The impedance of
the component under test is limited to a range from 20

 to

4.7k

. Below and above these values, the test pattern shows

only short-circuit or open-circuit. For the interpretation of the
displayed test pattern, these limits should always be borne
in mind. However, most electronic components can normally
be tested without any restriction.

Using the Component Tester

The instrument can be switched to component tester mode
by depressing 

COMP. TESTER

 pushbutton once. This is

completely independent from Yt or XY mode. In component
tester mode, only the X MAG. x10 LED may be lit  if it was
previously enabled. Then the X trace length is expanded 10
fold. All other LED‘s are unlit.

The following description concerns the operation without X
MAG. x10, with a horizontal trace length of approx. 8  div.
when the component tester input‘s are not connected. The
COMP. TESTER mode can be exited by depressing the COMP.
TESTER pushbutton again, setting the instrument back to
the previous mode and settings. The COMP. TESTER mode
can be saved and recalled as mentioned under SAVE/RECALL.

In component tester mode, the vertical preamplifier and the
timebase generator inoperative. A shortened horizontal trace
will be observed. It is not necessary to disconnect scope
input cables unless in-circuit measurements are to be carried
out. In the COMP. TESTER mode, the only controls which
can be operated are INTENS., FOCUS, X-POS. and X MAG.
x10. All other controls except the X MAG. x10 have no
influence on the test operation.

For the component connection, two simple test leads with
4mm 

Ø 

banana plugs, and with test prod, alligator clip or

sprung hook, are required. The test leads are connected to
the insulated socket and the adjacent ground socket beneath
the screen. The component can be connected to the test
leads either way round.

Test Procedure

Caution! Do not test any component in live circuitry -
remove all grounds, power and signals connected to
the component under test. Set up Component Tester
as stated above. Connect test leads across component
to be tested. Observe oscilloscope display.

Only discharged capacitors should be tested!

Test Pattern Displays

In the following typical test patterns show various
components under test.

Open circuit is indicated by a straight horizontal line.

Short circuit is shown by a straight vertical line.

Содержание HM304

Страница 1: ...tructions General Information 5 Symbols 5 Use of tilt handle 5 Safety 5 Operating conditions 5 EMC 6 Warranty 6 Maintenance 6 Protective Switch Off 6 Power supply 6 Type of signal voltage 7 Amplitude...

Страница 2: ...uction in the manual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection...

Страница 3: ...nzt durch 93 68 EWG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized...

Страница 4: ...35MHz 3dB Risetime 10ns Overshoot max 1 Deflection coefficients 14 calibrated steps from1mV div to20V div 1 2 5 sequence with variable 2 5 1 up to 50V div Accuracy in calibrated position 1mV div to 2m...

Страница 5: ...three conductor power cord with protective earthing conductor and a plug with earthing contact The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact T...

Страница 6: ...aintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largely certain that all signals are displayed with the accuracy on...

Страница 7: ...ms Veff have 2 83 times the potential difference in Vpp The relationship between the different voltage magnitudes can be seen from the following figure Voltage values of a sine curve Vrms effective va...

Страница 8: ...e base setting indicated by one of the TIME DIV LED s one or several signal periods or only a part of a period can be displayed The time coefficients are stated in s div when the red sec LED and the 0...

Страница 9: ...hes next to the edge are also not taken into account With very severe transient distortions the rise and fall time measurement has little meaning For amplifiers with approximately constant group delay...

Страница 10: ...oscilloscope by depressing the red POWER pushbutton The instrument will revert to its last used operating mode Except in the case of COMP TESTER mode where a trace appears on the screen if the INTENS...

Страница 11: ...should be adjusted The location of the low frequency compensation trimmer can be found in the probe information sheet Adjust the trimmer with the insulated screw driver provided until the tops of the...

Страница 12: ...channell I to channel II and vice versa after each sweep period In DUAL mode the internal trigger source can be switched over from channel I to channel II and vice versa if the TRIG pushbutton is dep...

Страница 13: ...compared with C short circuited then the test voltage leads the reference voltage and vice versa This applies only in the region up to 90 phase shift Therefore C should be sufficiently large and produ...

Страница 14: ...t when external trigger is used the trigger threshold can be stated as vertical display height in div through which the time base generator starts the display is stable and the trigger LED located in...

Страница 15: ...gnals than the DC coupling because the white noise in the trigger voltage is strongly suppressed So jitter or double triggering of complex signals is avoidable or at least reduced in particular with v...

Страница 16: ...er voltage may have a completely different form from the test signal voltage Triggering is even possible in certain limits with whole number multiples or fractions of the test frequency but only with...

Страница 17: ...following explanation assumes that the trace starts on the left vertical graticule line Photo 1 composite video signal MODE undelayed TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Depre...

Страница 18: ...ted in Dual mode under conditions where DUAL chopped mode is active this display mode is not switched off when time coefficients are being reduced 0 2ms div to 0 05 s div for signal expansion in DEL a...

Страница 19: ...also be made to integrated circuits All these components can be tested in and out of circuit The test principle is fascinatingly simple A built in generator delivers a sine voltage which is applied a...

Страница 20: ...ore testing of transistor amplification is not possible but testing of a single junction is easily and quickly possible Since the test voltage applied is only very low all sections of most semiconduct...

Страница 21: ...ould then be connected to the insulated COMP TESTER socket avoiding hum distortion of the test pattern Another way is a test pattern comparison to an identical circuit which is known to be operational...

Страница 22: ...Subject to change without notice 22...

Страница 23: ...of astigmatism A certain loss of marginal sharpness of the CRT is unavoidable this is due to the manufacturing process of the CRT Symmetry and Drift of the Vertical Amplifier Both of these characteri...

Страница 24: ...eck of the mono channel display is unnecessary it is contained indirectly in the tests above stated Triggering Checks The internal trigger threshold is important as it determines the display height fr...

Страница 25: ...and also the influence of the earths magnetic field which is dependent on the instruments North orientation are corrected by means of the TR potentiometer In general the trace rotation range is asymme...

Страница 26: ...hanged Generally max halving or doubling of this resistance value should be sufficient A too small trigger threshold cause double triggering or premature trigger action due to interference pulses or r...

Страница 27: ...R ckgabe Beschreibung PC Scope Scope PC ID ID Daten CR LF data consits of instrument type manufacturer CR R CR LF remote status and baud rate acceptance TRSTA TRSTA b CR LF query for trigger status d...

Страница 28: ...nter 0 13 CH2 GND AC INV2 ON VALUE Counter 0 13 mv DIV 0000 20V DIV 1101 MODE CT XY A TR CHOP ADD 0 TR SOURCE 00 Y1 01 Y2 1x EXT TB1 x10 0 0 TB A TIME Counter 1 26 50ns DIV 00 bis 0 5s DIV 15hex TB2 0...

Страница 29: ...cy ranges AC 10Hz 100MHz DC 0Hz 100MHz HF 1 5kHz 100MHz LF 0Hz 1 5 kHz TV L to trigger on line sync pulses TV F to trigger on separated frame sync pulses Select SLOPE for the leading slope Sync pulse...

Страница 30: ...control for trace sharpness mechanical knob TR Trace rotation mechanical To align trace with horzontal field potentiometer graticule line Compensates adjustment with influence of Earth s magnetic scr...

Страница 31: ...triggering Not available in combination with ext triggering XY or COMP TESTER modes INPUT CH II Channel II signal input BNC connector Input impedance 1M II 20pF AC DC Selects input coupling of CH II...

Отзывы: