P2006:
DC-Triggering CH I
– Locate and identify P2006 on YP-Board.
– Connect a 50 kHz sine wave signal to input CH I.
– Set time base to 5 μs/div.
– Set generator amplitude for a display height of 8 divi-
sions.
– Check that input coupling CH I is set to AC.
– Set trigger coupling to AC.
– Press AT/NM (Automatic/Norm) pushbutton (IN position).
– Set LEVEL knob for a trigger point at the centre of the signal
slope and note the position (reference).
– Set trigger coupling to DC.
– Adjust P2006 so that the trigger point (trace start) is in the
same position as under AC coupling condition.
Reference
Incorrect
Correct
Incorrect
36
P2008:
DC-Triggering CH II
– Locate and identify P2008 on YP-Board.
– Connect a 50 kHz sine wave signal to input CH I.
– Set time base to 5 μs/div.
– Set generator amplitude for a display height of 8 divi-
sions.
– Check that input coupling CH I is set to AC.
– Set trigger coupling to AC.
– Press AT/NM (Automatic/Norm) pushbutton (IN position).
– Set LEVEL knob for a trigger point at the centre of the signal
slope and note the position (reference).
– Set trigger coupling to DC.
– Adjust P2008 so that the trigger point (trace start) is in the
same position as under AC coupling condition.
Reference
Incorrect
Correct
Incorrect
P2008
P2011 P2012
P2006
P2007
P2005
P2002
P2001
C2007
C2010
C2009
C2011
P2003
C2107
C2108
C2103
C2110
P2016
P2014
P2015
P2010
P2009
YP-Board
P2008
P2011 P2012
P2006
P2007
P2005
P2002
P2001
C2007
C2010
C2009
C2011
P2003
C2107
C2108
C2103
C2110
P2016
P2014
P2015
P2010
P2009
YP-Board
A d j u s t m e n t P r o c e d e r e H M 3 0 3 - 6